Improving fat saturation robustness in outer extremity MRI with a local shim coil insert
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30474 ↗At a glance
Journal article in Magnetic Resonance in Medicine, 2025. Written by Ziyang Long and 9 co-authors. Institution: Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA. Major MEDLINE topics: Magnetic Resonance Imaging, Adipose Tissue, Image Enhancement. Cited 5 times so far (Crossref); 35 references. The full text is freely available in PubMed Central. Funded by: NINDS NIH HHS, NHLBI NIH HHS, National Heart, Lung, and Blood Institute.
Abstract (original language)
Purpose
High‐quality fat suppression is essential for various MRI applications. In musculoskeletal imaging, poor fat suppression caused by severe B 0 inhomogeneity can obscure important lesions, potentially leading to inaccurate diagnoses. This problem is particularly exacerbated in off‐isocenter imaging, where conventional shimming using second‐order spherical harmonic shim coils often proves inadequate due to elevated B 0 inhomogeneity. To address this challenge, we configured a simple local shim insert to provide additional localized B 0 shimming for off‐isocenter regions, offering a practical hardware solution.
Methods
We designed and constructed a seven‐channel shim coil and evaluated its performance in comparison to conventional second‐order spherical harmonic shimming within a targeted volume near the scanner bore. The coil was tested with both phantom and in vivo studies using a clinical 3 T MRI scanner.
Results
The improved B 0 homogeneity achieved with the local shim coil significantly enhanced fat saturation (fat–sat) uniformity across the imaged volumes. This improvement was particularly beneficial for areas far from the scanner isocenter, where B 0 inhomogeneity is most severe. Our results indicated a 40% reduction in RMS error of the B 0 field for elbow imaging and 35% for hand imaging, highlighting substantial improvements in B 0 field homogeneity. Additionally, the image quality score increased by 1 point for both hand and elbow images, reflecting enhanced fat–sat quality.
Conclusion
The simple local shim insert we configured improves fat–sat capability in both hand and elbow imaging. It offers the potential for improving off‐isocenter musculoskeletal MRI.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Biomedical Imaging Research Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA.
- Department of Radiology, Weill Cornell Medicine, New York, New York, USA.
- Department of Bioengineering, University of California, Los Angeles, Los Angeles, California, USA.
- Department of Imaging, Cedars-Sinai Medical Center, Los Angeles, California, USA.
ORCID profiles: Ziyang Long, Archana Vadiraj Malagi, Fardad Michael Serry, Hui Han
Funding
- NINDS NIH HHS
- NHLBI NIH HHS
- National Heart, Lung, and Blood Institute
- National Institute of Neurological Disorders and Stroke
Cite
APA
Long, Z., Yang, H. J., Binesh, N., Malagi, A. V., Shang, Y., Huang, L. T., Zepeda, J., Serry, F. M., Li, D., & Han, H. (2025). Improving fat saturation robustness in outer extremity MRI with a local shim coil insert. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30474
Vancouver
Long Z, Yang HJ, Binesh N, Malagi AV, Shang Y, Huang LT, et al. Improving fat saturation robustness in outer extremity MRI with a local shim coil insert. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30474
BibTeX
@article{Long2025,
author = {Long, Ziyang and Yang, Hsin-Jung and Binesh, Nader and Malagi, Archana Vadiraj and Shang, Yun and Huang, Li-Ting and Zepeda, Jeremy and Serry, Fardad Michael and Li, Debiao and Han, Hui},
title = {Improving fat saturation robustness in outer extremity MRI with a local shim coil insert},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30474}
}
EUVIMED