Prospective compensation of second‐order concomitant fields in a high‐performance gradient system using a second‐order harmonic shim coil
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70094 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Afis Ajala und 2 weiteren Autor:innen. Beteiligte Einrichtung: GE HealthCare, Technology & Innovation Center, Niskayuna, New York, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain, Image Processing, Computer-Assisted, Image Enhancement. Die Arbeit nennt 39 Literaturangaben. Gefördert von: CDMRP, Congressionally Directed Medical Research Programs.
Abstract (Originalsprache)
Purpose
The use of high‐performance gradient coils results in stronger spatially dependent second‐order concomitant magnetic fields, which can lead to signal dropout, blurring artifacts and phase errors that become more significant at locations farther from the gradient isocenter. A correction coil‐based method for prospectively compensating second‐order concomitant fields in higher‐performance gradient systems is described.
Methods
An insertable, axially symmetric second‐order field coil to prospectively correct for second‐order concomitant field‐induced phase errors on a high‐performance head‐only gradient system at 3.0T was developed. The efficacy of the second‐order concomitant‐field correction was demonstrated in phantom and healthy volunteer scans using 2D phase contrast (PC) and spiral gradient echo (GRE) imaging.
Results
By employing the correction coil, there was a significant reduction in second‐order concomitant field‐induced blurring in the 2D spiral images, and reduction in phase errors and signal degradation in the GRE PC images. In the single‐sided PC, the z‐ and radially‐directed second‐order concomitant phase accrued in the coronal and axial PC acquisition was reduced by 100% and 83%, respectively. Signal enhancement up to 968.9% was obtained in the two‐sided PC acquisitions. In spiral GRE images, blurring was reduced by ˜40.2% at 60 mm from the gradient isocenter in a phantom. Correspondingly, the reduction in concomitant field‐induced blurring in in‐vivo spiral GRE images was noted with the correction coil.
Conclusions
The described second‐order correction coil insert prospectively compensates erroneous phase accruals due to second‐order concomitant fields on a high‐performance gradient system at the source, complementing or replacing software corrections/compensations during image reconstruction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- GE HealthCare, Technology & Innovation Center, Niskayuna, New York, USA.
ORCID-Profile: Afis Ajala
Förderung
- CDMRP
- Congressionally Directed Medical Research Programs
Zitieren
APA
Ajala, A., Foo, T. K. F., & Lee, S. K. (2026). Prospective compensation of second‐order concomitant fields in a high‐performance gradient system using a second‐order harmonic shim coil. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70094
Vancouver
Ajala A, Foo TKF, Lee SK. Prospective compensation of second‐order concomitant fields in a high‐performance gradient system using a second‐order harmonic shim coil. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70094
BibTeX
@article{Ajala2026,
author = {Ajala, Afis and Foo, Thomas K F and Lee, Seung-Kyun},
title = {Prospective compensation of second‐order concomitant fields in a high‐performance gradient system using a second‐order harmonic shim coil},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70094}
}
EUVIMED