MR sequence design to account for nonideal gradient performance
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70093 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Daniel J West und 6 weiteren Autor:innen. Beteiligte Einrichtung: Imaging Physics & Engineering Research Department, School of Biomedical Engineering & Imaging Sciences, King's College London, London, UK. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Image Enhancement. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Wellcome Trust, National Institute for Health Research (NIHR), Wellcome/EPSRC Centre for Medical Engineering.
Abstract (Originalsprache)
Purpose
MRI systems are traditionally engineered to produce close to idealized performance, enabling a simplified pulse sequence design philosophy. An example of this is control of eddy currents produced by gradient fields; usually these are compensated by pre‐emphasizing demanded waveforms. This process typically happens invisibly to the pulse sequence designer, allowing them to assume achieved gradient waveforms will be as desired. Although convenient, this requires system specifications exposed to the end user to be substantially down‐rated, as pre‐emphasis adds an extra overhead to the waveforms. This strategy is undesirable for lower performance or resource‐limited hardware. Instead, we propose an optimization‐based method to design precompensated gradient waveforms that (i) explicitly respect hardware constraints and (ii) improve imaging performance by correcting k‐space samples directly.
Methods
Gradient waveforms are numerically optimized by including a model for system imperfections. This is investigated in simulation using an exponential eddy current model, then experimentally using an empirical gradient system transfer function on a 7T MRI system.
Results
Our proposed method discovers solutions that produce negligible reconstruction errors while satisfying gradient system limits, even when classic pre‐emphasis produces infeasible results. Substantial reduction in ghosting artifacts from echo‐planar imaging was observed, including an average reduction of 77% in ghost amplitude in phantoms.
Conclusions
This work demonstrates numerical optimization of gradient waveforms, yielding substantially improved image quality when given a model for system imperfections. Although the method as implemented has limited flexibility, it could enable more efficient hardware use and may prove particularly important for maximizing performance of lower‐cost systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Imaging Physics & Engineering Research Department, School of Biomedical Engineering & Imaging Sciences, King's College London, London, UK.
- Max-Planck-Institut fur Biologische Kybernetik, Magnetic Resonance Center, Tübingen, Germany.
- Institute of Neuroradiology, Erlangen University Hospital Institute of Neuroradiology, Erlangen, Germany.
- Department of Artificial Intelligence in Biomedical Engineering, Friedrich-Alexander-Universität Erlangen-Nurnberg, Erlangen, Germany.
- Early Life Imaging Research Department, School of Biomedical Engineering & Imaging Sciences, King's College London, London, UK.
ORCID-Profile: Daniel J West, Felix Glang, Jonathan Endres, David Leitão, Moritz Zaiss, Joseph V Hajnal, Shaihan J Malik
Förderung
- Wellcome Trust
- National Institute for Health Research (NIHR)
- Wellcome/EPSRC Centre for Medical Engineering
- National Institute for Health and Care Research
- Wellcome EPSRC Centre for Medical Engineering
Zitieren
APA
West, D. J., Glang, F., Endres, J., Leitão, D., Zaiss, M., Hajnal, J. V., & Malik, S. J. (2026). MR sequence design to account for nonideal gradient performance. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70093
Vancouver
West DJ, Glang F, Endres J, Leitão D, Zaiss M, Hajnal JV, et al. MR sequence design to account for nonideal gradient performance. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70093
BibTeX
@article{West2026,
author = {West, Daniel J and Glang, Felix and Endres, Jonathan and Leitão, David and Zaiss, Moritz and Hajnal, Joseph V and Malik, Shaihan J},
title = {MR sequence design to account for nonideal gradient performance},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70093}
}
EUVIMED