EUVIMEDOpdag EUVIMED
Open access

Water phase transition and signal nulling in 3D dual‐echo adiabatic inversion‐recovery UTE ( IR ‐ UTE ) imaging of myelin

Jiyo S Athertya iD, Soo Hyun Shin iD, Bhavsimran Singh Malhi, James Lo, Sam Sedaghat, Hyungseok Jang iD, Yajun Ma iD, Jiang Du

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30243 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, 2024. Written by Jiyo S Athertya and 7 co-authors. Institution: Department of Radiology, University of California, San Diego, California, USA. Major MEDLINE topics: Myelin Sheath, Magnetic Resonance Imaging, Imaging, Three-Dimensional. Cited 1 times so far (Crossref); 30 references. The full text is freely available in PubMed Central. Funded by: CSRD VA, NIA NIH HHS, Clinical Science Research and Development.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Purpose

The semisolid myelin sheath has very fast transverse relaxation and is invisible to conventional MRI sequences. UTE sequences can detect signal from myelin. The major challenge is the concurrent detection of various water components.

Methods

The inversion recovery (IR)–based UTE (IR‐UTE) sequence employs an adiabatic inversion pulse to invert and suppress water magnetizations. TI plays a key role in water suppression, with negative water magnetizations (negative phase) before the null point and positive water magnetizations (positive phase) after the null point. A series of dual‐echo IR‐UTE images were acquired with different TIs to detect water phase transition. The effects of TR in phase transition and water suppression were also investigated using a relatively long TR of 500 ms and a short TR of 106 ms. The water phase transition in dual‐echo IR‐UTE imaging of myelin was investigated in five ex vivo and five in vivo human brains.

Results

An apparent phase transition was observed in the second echo at the water signal null point, where the myelin signal was selectively detected by the UTE data acquisition at the optimal TI. The water phase transition point varied significantly across the brain when the long TR of 500 ms was used, whereas the convergence of TIs was observed when the short TR of 106 ms was used.

Conclusion

The results suggest that the IR‐UTE sequence with a short TR allows uniform inversion and nulling of water magnetizations, thereby providing volumetric imaging of myelin.

Subjects

MeSH terms (MEDLINE)

Myelin SheathMagnetic Resonance ImagingImaging, Three-DimensionalHumansBrainWaterAlgorithmsPhase TransitionBody WaterImage Interpretation, Computer-AssistedMaleFemaleAdultReproducibility of ResultsImage Enhancement

Author keywords

MRIUTEmyelinphase transitionwhite matter

Substances

Water

Authors and institutions

  • Department of Radiology, University of California, San Diego, California, USA.
  • Department of Bioengineering, University of California, San Diego, California, USA.
  • Department of Diagnostic and Interventional Radiology, University Hospital Heidelberg, Heidelberg, Germany.
  • Department of Radiology, University of California, Davis, California, USA.
  • Radiology Service, Veterans Affairs San Diego Healthcare System, San Diego, California, USA.

ORCID profiles: Jiyo S Athertya, Soo Hyun Shin, Hyungseok Jang, Yajun Ma

Funding

  • CSRD VA
  • NIA NIH HHS
  • Clinical Science Research and Development
  • National Institutes of Health

Cite

APA

Athertya, J. S., Shin, S. H., Malhi, B. S., Lo, J., Sedaghat, S., Jang, H., Ma, Y., & Du, J. (2024). Water phase transition and signal nulling in 3D dual‐echo adiabatic inversion‐recovery UTE ( IR ‐ UTE ) imaging of myelin. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30243

Vancouver

Athertya JS, Shin SH, Malhi BS, Lo J, Sedaghat S, Jang H, et al. Water phase transition and signal nulling in 3D dual‐echo adiabatic inversion‐recovery UTE ( IR ‐ UTE ) imaging of myelin. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30243

BibTeX
@article{Athertya2024,
  author = {Athertya, Jiyo S and Shin, Soo Hyun and Malhi, Bhavsimran Singh and Lo, James and Sedaghat, Sam and Jang, Hyungseok and Ma, Yajun and Du, Jiang},
  title = {Water phase transition and signal nulling in 3D dual‐echo adiabatic inversion‐recovery UTE ( IR ‐ UTE ) imaging of myelin},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30243}
}
Start for free →