Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Ziwen Ke und 7 weiteren Autor:innen. Beteiligte Einrichtung: Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, Urbana, Illinois, USA. Hauptthemen laut MEDLINE: Brain, Imaging, Three-Dimensional, Magnetic Resonance Imaging. Die Arbeit nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Purpose
To develop an effective method for correcting head motion in high‐resolution, non‐water‐suppressed MRSI.
Methods
MRSI scans are susceptible to subject motion due to the long data acquisition time required for sufficient spatial‐spectral encodings. The problem is more serious in non‐water‐suppressed MRSI experiments since motion artifacts in the water and lipid signals make their removal even more challenging. To address this problem, we propose a novel motion correction method that detects and discards motion‐corrupted k‐space data using TR‐wise linear navigators. The discarded data are replaced with reconstructed data obtained by reformulating motion correction as a missing data reconstruction problem using sensitivity encodings. Extrinsic priors for water and lipid signals and intrinsic priors for metabolite signals are incorporated in the motion correction process. We evaluated its performance on a spectroscopic phantom and three human groups: (1) five healthy adults performing three different voluntary motion patterns; (2) a healthy child; and (3) a cerebral hemorrhage patient with involuntary movements.
Results
In phantom and healthy subjects, the proposed method produced high‐quality water images and metabolite maps that closely matched motion‐free references, with 10%–20% quantitative gains in image/map quality (higher PSNR/SSIM, lower NRMSE). In the child and patient data, motion artifacts were noticeably reduced, with 20%–30% reductions in NAA linewidth and fitting error in the child, and ˜50% reductions in coefficient‐of‐variation in the patient.
Conclusion
An effective motion‐correction technique has been developed for high‐resolution non‐water‐suppressed MRSI. This method has the potential to significantly enhance the robustness and clinical applicability of non‐water‐suppressed MRSI.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Beckman Institute for Advanced Science and Technology, University of Illinois at Urbana-Champaign, Urbana, Illinois, USA.
- National Engineering Research Center of Advanced Magnetic Resonance Technologies for Diagnosis and Therapy, School of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai, China.
- Siemens Medical Solutions USA, Inc., St. Louis, Missouri, USA.
- Department of Bioengineering, University of Illinois at Urbana-Champaign, Urbana, Illinois, USA.
- Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, Illinois, USA.
ORCID-Profile: Ziwen Ke, Yibo Zhao, Rong Guo, Yudu Li, Wen Jin, Yao Li
Förderung
- NIBIB NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Ke, Z., Zhao, Y., Guo, R., Li, Y., Jin, W., Zhuang, H., Li, Y., & Liang, Z. P. (2026). Motion Correction in High‐Resolution 3D Brain MRSI Without Water and Lipid Suppression. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70128
Vancouver
Ke Z, Zhao Y, Guo R, Li Y, Jin W, Zhuang H, et al. Motion Correction in High‐Resolution 3D Brain MRSI Without Water and Lipid Suppression. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70128
BibTeX
@article{Ke2026,
author = {Ke, Ziwen and Zhao, Yibo and Guo, Rong and Li, Yudu and Jin, Wen and Zhuang, Huixiang and Li, Yao and Liang, Zhi-Pei},
title = {Motion Correction in High‐Resolution 3D Brain MRSI Without Water and Lipid Suppression},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70128}
}
EUVIMED