Water phase transition and signal nulling in 3D dual‐echo adiabatic inversion‐recovery UTE ( IR ‐ UTE ) imaging of myelin
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30243 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Jiyo S Athertya und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, University of California, San Diego, California, USA. Hauptthemen laut MEDLINE: Myelin Sheath, Magnetic Resonance Imaging, Imaging, Three-Dimensional. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: CSRD VA, NIA NIH HHS, Clinical Science Research and Development.
Abstract (Originalsprache)
Purpose
The semisolid myelin sheath has very fast transverse relaxation and is invisible to conventional MRI sequences. UTE sequences can detect signal from myelin. The major challenge is the concurrent detection of various water components.
Methods
The inversion recovery (IR)–based UTE (IR‐UTE) sequence employs an adiabatic inversion pulse to invert and suppress water magnetizations. TI plays a key role in water suppression, with negative water magnetizations (negative phase) before the null point and positive water magnetizations (positive phase) after the null point. A series of dual‐echo IR‐UTE images were acquired with different TIs to detect water phase transition. The effects of TR in phase transition and water suppression were also investigated using a relatively long TR of 500 ms and a short TR of 106 ms. The water phase transition in dual‐echo IR‐UTE imaging of myelin was investigated in five ex vivo and five in vivo human brains.
Results
An apparent phase transition was observed in the second echo at the water signal null point, where the myelin signal was selectively detected by the UTE data acquisition at the optimal TI. The water phase transition point varied significantly across the brain when the long TR of 500 ms was used, whereas the convergence of TIs was observed when the short TR of 106 ms was used.
Conclusion
The results suggest that the IR‐UTE sequence with a short TR allows uniform inversion and nulling of water magnetizations, thereby providing volumetric imaging of myelin.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, University of California, San Diego, California, USA.
- Department of Bioengineering, University of California, San Diego, California, USA.
- Department of Diagnostic and Interventional Radiology, University Hospital Heidelberg, Heidelberg, Germany.
- Department of Radiology, University of California, Davis, California, USA.
- Radiology Service, Veterans Affairs San Diego Healthcare System, San Diego, California, USA.
ORCID-Profile: Jiyo S Athertya, Soo Hyun Shin, Hyungseok Jang, Yajun Ma
Förderung
- CSRD VA
- NIA NIH HHS
- Clinical Science Research and Development
- National Institutes of Health
Zitieren
APA
Athertya, J. S., Shin, S. H., Malhi, B. S., Lo, J., Sedaghat, S., Jang, H., Ma, Y., & Du, J. (2024). Water phase transition and signal nulling in 3D dual‐echo adiabatic inversion‐recovery UTE ( IR ‐ UTE ) imaging of myelin. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30243
Vancouver
Athertya JS, Shin SH, Malhi BS, Lo J, Sedaghat S, Jang H, et al. Water phase transition and signal nulling in 3D dual‐echo adiabatic inversion‐recovery UTE ( IR ‐ UTE ) imaging of myelin. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30243
BibTeX
@article{Athertya2024,
author = {Athertya, Jiyo S and Shin, Soo Hyun and Malhi, Bhavsimran Singh and Lo, James and Sedaghat, Sam and Jang, Hyungseok and Ma, Yajun and Du, Jiang},
title = {Water phase transition and signal nulling in 3D dual‐echo adiabatic inversion‐recovery UTE ( IR ‐ UTE ) imaging of myelin},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30243}
}
EUVIMED