Enhanced analysis of gating latency in 0.35T MR‐linac through innovative time synchronization of a motion phantom and plastic scintillation detector
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70116 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Mateb Al Khalifa und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Algorithms, Particle Accelerators, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This study aims to evaluate how different gantry angles, breathing rates (BPM), cine image speeds, and tracking algorithms affect beam on/off latency and the subsequent impact on target dose for a 0.35T MR‐Linac with a 6 MV FFF beam. The investigation incorporates an image‐based MRI4D modus QA motion phantom (MQA) and a measurement‐based plastic scintillation detector (PSD).
Methods
The MQA's target was customized with an insertion for a 1 mm PSD from BluePhysics. Both the PSD and the MQA were simultaneously synchronized to the Linac to capture latency signals. A plan was created in the ViewRay TPS to deliver dose to the target at three gantry angles (0°, 120°, and 240°). Each gantry angle was evaluated at three breathing rates (10, 12, and 15 BPM). The study also examined two imaging speeds (4 and 8 FPS) and four tracking algorithms.
Results
Across all configurations at 4 FPS, the overall mean beam‐on latency was 0.339 ± 0.06 s from the PSD and 0.318 ± 0.06 s from the MQA, whereas at 8 FPS it was 0.630 ± 0.07 s (PSD) and 0.609 ± 0.07 s (MQA). Conversely, the overall mean beam‐off latency at 4 FPS was 0.153 ± 0.03 s (PSD) and 0.124 ± 0.03 s (MQA), while at 8 FPS it was 0.121 ± 0.06 s (PSD) and 0.205 ± 0.04 s (MQA). The overall mean difference between gating and non‐gating doses was an increase of 12.050 ± 9.2 cGy at 4 FPS and 14.044 ± 7.4 cGy at 8 FPS.
Conclusion
This comprehensive study underscores the significant influence of gantry angle, breathing rate, cine imaging speed, and tracking algorithms on latency and dose delivery accuracy in a 0.35T MR‐Linac.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA.
ORCID-Profile: Mateb Al Khalifa, Siyong Kim, William Y Song
Zitieren
APA
Khalifa, M. A., Ma, T., Aljuaid, H., Kim, S., & Song, W. Y. (2025). Enhanced analysis of gating latency in 0.35T MR‐linac through innovative time synchronization of a motion phantom and plastic scintillation detector. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70116
Vancouver
Khalifa MA, Ma T, Aljuaid H, Kim S, Song WY. Enhanced analysis of gating latency in 0.35T MR‐linac through innovative time synchronization of a motion phantom and plastic scintillation detector. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70116
BibTeX
@article{Khalifa2025,
author = {Khalifa, Mateb Al and Ma, Tianjun and Aljuaid, Haya and Kim, Siyong and Song, William Y},
title = {Enhanced analysis of gating latency in 0.35T MR‐linac through innovative time synchronization of a motion phantom and plastic scintillation detector},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70116}
}
EUVIMED