Commissioning and implementation of a pencil‐beam algorithm with a Lorentz correction as a secondary dose calculation algorithm for an Elekta Unity 1.5T MR linear accelerator
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.14590 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Sameer Taneja und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, New York University Langone Medical Center, New York, New York, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Particle Accelerators, Algorithms, Phantoms, Imaging. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To commission a beam model in ClearCalc (Radformation Inc.) for use as a secondary dose calculation algorithm and to implement its use into an adaptive workflow for an MR‐linear accelerator.
Methods
A beam model was developed using commissioning data for an Elekta Unity MR‐linear accelerator and entered into ClearCalc. The beam model consisted of absolute dose calculation settings, output factors, percent depth‐dose (PDD) curves, mutli‐leaf collimator (MLC) transmission and dose leaf gap error, and cryostat corrections. Beam profiles were hard‐coded by the manufacturer into the beam model and were compared with Monaco‐derived profiles. The beam model was tested by comparing point doses in a homogenous phantom obtained through measurements using an ionization chamber in water, Monaco, and ClearCalc for various field sizes, source‐surface distances (SSDs), and point locations. Additional testing including point dose verification for test plans using a heterogeneous phantom and patient plans. Post clinical implementation, performance of ClearCalc was evaluated for the first 41 patients treated, which included 215 adaptive plans.
Results
PDDs generated using ClearCalc fell within 1.2% of measurements. Field profile comparison between ClearCalc and Monaco showed an average pass rate of 98% using a 3%/3 mm gamma criteria. Measured cryostat corrections used in the beam model showed a maximum deviation from unity of 1.4%. Point dose and field monitor units (MUs) comparisons in a homogenous phantom ( N = 22), heterogeneous phantoms ( N = 22), and patient plans ( N = 57) all passed with a threshold of 5%/5MU. Clinically, ClearCalc was implemented as a physics check post adaptive planning completed prior to beam delivery. Point dose and field MUs showed good agreement at a 5%/5MU threshold for prostate stereotactic body radiation therapy (SBRT), pelvic lymph nodes, rectum, and prostate and lymph node plans.
Discussion
This work demonstrated commissioning and clinical implementation of ClearCalc into an adaptive planning workflow. No primary or adaptive plan failures were reported with proper beam model testing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, New York University Langone Medical Center, New York, New York, USA.
- Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut, USA.
Zitieren
APA
Taneja, S., Wang, H., Barbee, D. L., Galavis, P., Sosa, M. S., Byun, D., Zelefsky, M., & Chen, T. (2025). Commissioning and implementation of a pencil‐beam algorithm with a Lorentz correction as a secondary dose calculation algorithm for an Elekta Unity 1.5T MR linear accelerator. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14590
Vancouver
Taneja S, Wang H, Barbee DL, Galavis P, Sosa MS, Byun D, et al. Commissioning and implementation of a pencil‐beam algorithm with a Lorentz correction as a secondary dose calculation algorithm for an Elekta Unity 1.5T MR linear accelerator. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.14590
BibTeX
@article{Taneja2025,
author = {Taneja, Sameer and Wang, Hesheng and Barbee, David L and Galavis, Paulina and Sosa, Mario Serrano and Byun, David and Zelefsky, Michael and Chen, Ting},
title = {Commissioning and implementation of a pencil‐beam algorithm with a Lorentz correction as a secondary dose calculation algorithm for an Elekta Unity 1.5T MR linear accelerator},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.14590}
}
EUVIMED