Fast and accurate measurement of small field dosimetry using a novel scintillation detector

Yiding Han iD, Jingzhu Xu, Yao Hao iD, Baozhou Sun

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70406 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Yiding Han und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Baylor College of Medicine, Houston, Texas, USA. Hauptthemen laut MEDLINE: Scintillation Counting, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Radiometry. Die Arbeit nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The most used instruments for small‐field dosimetry have notable limitations, including the need for correction of output factors, limited scanning speeds, and challenges in alignment for percentage depth dose (PDD) measurements, particularly for extremely small fields. However, plastic scintillation detectors (PSDs) are an attractive alternative for small‐field dosimetry due to their correction‐free nature, linear dose response, and fast response time.

Purpose

This study evaluates the robustness and accuracy of the dosimetric measurements using a new water‐equivalent PSD in small‐field dosimetry. The study also aims to report accurate measurements of output factors, profiles, and an indirect method for measuring PDD in small fields, with a scanning time that is 5 to 10 times faster than traditional methods.

Method

PDDs, profiles, and output factors were measured on a Varian TrueBeam 6XFFF photon beam for the field sizes of 0.5 × 0.5 cm 2 , 1 × 1 cm 2 , 2 × 2 cm 2 , 3 × 3 cm 2 , and 4 × 4 cm 2 using a new PSD from blue physics (BP‐PSD). These measurements were compared with those obtained using a well‐established PSD (Standard Imaging W2), micro‐diamond (TN60019, PTW‐Freiburg, Germany), and micro‐silicon detectors (TN60023, PTW‐Freiburg, Germany). Owing to its fast response, the BP‐PSD enabled the collection of beam profiles at 31 depths, which were used to derive the PDD while avoiding detector misalignment along the beam path. Data was collected in a water tank controlled by the PTW BeamScan software. The pulse‐by‐pulse raw data from BP‐PSD were converted to respective dosimetry data using in‐house software.

Result

The BP‐PSD demonstrated excellent agreement with other detectors for small‐field output factors (FOFs), with a maximum variation of 1.6%. The BP‐PSD also showed strong agreement in PDD measurements with an ion chamber (TN31013) for both 3 × 3 cm 2 and 10 × 10 cm 2 field sizes, achieving a 98% gamma passing rate (gamma criteria: 1 mm, 3%). For the profile measurements, the BP‐PSD showed consistency with both the micro‐diamond and micro‐silicon diode detectors, with less than 1% variation in measured penumbra length. At a 3 × 3 cm 2 field size, the measured penumbra length (4 mm) agreed with previously published data (3.86–4.2 mm). Additionally, for an extremely small field size (0.5 × 0.5 cm 2 ), the indirect PDD measurements derived from profiles showed significant improvement compared to the direct measurement without beam inclination correction.

Conclusion

The BP‐PSD has proven to be a robust and reliable detector for small‐field dosimetry. It shows excellent agreement with other detectors in measuring small FOFs, profiles, and PDDs, while achieving significantly faster scanning speeds in a water tank.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Scintillation CountingPhantoms, ImagingRadiotherapy Planning, Computer-AssistedRadiometryHumansRadiotherapy DosagePhotonsParticle AcceleratorsRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

photon beamplastic scintillation detectorsmall‐field dosimetry

Autor:innen und Institutionen

  • Department of Radiation Oncology, Baylor College of Medicine, Houston, Texas, USA.
  • Washington University in St. Louis, St. Louis, Missouri, USA.

ORCID-Profile: Yiding Han, Yao Hao

Zitieren

APA

Han, Y., Xu, J., Hao, Y., & Sun, B. (2025). Fast and accurate measurement of small field dosimetry using a novel scintillation detector. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70406

Vancouver

Han Y, Xu J, Hao Y, Sun B. Fast and accurate measurement of small field dosimetry using a novel scintillation detector. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70406

BibTeX
@article{Han2025,
  author = {Han, Yiding and Xu, Jingzhu and Hao, Yao and Sun, Baozhou},
  title = {Fast and accurate measurement of small field dosimetry using a novel scintillation detector},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70406}
}
Kostenlos starten →