Ultra‐fast dosimetric data collection with a commercial plastic scintillation detector in an MR‐linac
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70460 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Carlos Ferrer und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics and Radiation Protection, La Paz University Hospital, Paseo de la Castellana 261, Madrid, Spain. Hauptthemen laut MEDLINE: Scintillation Counting, Plastics, Particle Accelerators, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Plastic scintillation detectors (PSD) are widely used for detecting and measuring ionizing radiation. These detectors are versatile, with high efficiency, fast response and the ability to provide real‐time measurements.
Purpose
Evaluate the suitability of Blue Physics PSD (BP‐PSD) for performing ultra‐fast dosimetric commissioning measurements with high accuracy and precision in a very short time.
Methods
Ultra‐fast measurements were performed in water using a BP‐PSD on an Elekta Unity MR‐linac. Percentage depth doses (PDD) and profiles at different depths were measured at two movement velocities, 10 mm/s and 20 mm/s, for field sizes ranging from 10 × 10 cm 2 to 1 × 1 cm 2 . Gamma analysis was conducted to compare these measurements with those obtained during machine commissioning using a PTW Semiflex 3D ionization chamber (for PDD) and a PTW micro‐Diamond detector (for PDD and profiles). Gamma criteria of 2%/2 mm and 1%/1 mm dose difference/distance to agreement were studied, alongside field size, penumbra, and measurement time.
Results
All PDD and profile gamma passing rates were 100% at 2%/2 mm. At the stricter 1%/1 mm criteria, all PDD showed a passing rate above 96.97% for both velocities, with most of the profiles exceeding 95% at 10 mm/s and 90% at 20 mm/s. Gamma analysis results were superior for smaller fields (1 × 1 cm 2 and 2 × 2 cm 2 ) and generally better at 10 mm/s. On average, the penumbra measurements obtained with the PSD were greater than those achieved with the micro‐Diamond detector. Measurement times were found to be between 7 and 14 times shorter for PDD, and between 5 and 9 times shorter for profiles at speeds of 10 mm/s and 20 mm/s, respectively.
Conclusions
Ultra‐fast measurements using the Blue Physics PSD are suitable for acquiring dosimetric commissioning data with high accuracy and precision, and can be performed in a much shorter timeframe than with commonly used detectors.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Medical Physics and Radiation Protection, La Paz University Hospital, Paseo de la Castellana 261, Madrid, Spain.
- Blue Physics LLC, 19145 long lake ranch blvd, Lutz, Florida, USA.
ORCID-Profile: Carlos Ferrer
Zitieren
APA
Ferrer, C., Huertas, C., Feijoo, M., Cardin, A., & Sáez, M. (2026). Ultra‐fast dosimetric data collection with a commercial plastic scintillation detector in an MR‐linac. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70460
Vancouver
Ferrer C, Huertas C, Feijoo M, Cardin A, Sáez M. Ultra‐fast dosimetric data collection with a commercial plastic scintillation detector in an MR‐linac. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70460
BibTeX
@article{Ferrer2026,
author = {Ferrer, Carlos and Huertas, Concepción and Feijoo, Marcos and Cardin, Alessandro and Sáez, Moisés},
title = {Ultra‐fast dosimetric data collection with a commercial plastic scintillation detector in an MR‐linac},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70460}
}
EUVIMED