Ultra‐fast dosimetric data collection with a commercial plastic scintillation detector in an MR‐linac

Carlos Ferrer iD, Concepción Huertas, Marcos Feijoo, Alessandro Cardin, Moisés Sáez

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70460 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Carlos Ferrer und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics and Radiation Protection, La Paz University Hospital, Paseo de la Castellana 261, Madrid, Spain. Hauptthemen laut MEDLINE: Scintillation Counting, Plastics, Particle Accelerators, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Plastic scintillation detectors (PSD) are widely used for detecting and measuring ionizing radiation. These detectors are versatile, with high efficiency, fast response and the ability to provide real‐time measurements.

Purpose

Evaluate the suitability of Blue Physics PSD (BP‐PSD) for performing ultra‐fast dosimetric commissioning measurements with high accuracy and precision in a very short time.

Methods

Ultra‐fast measurements were performed in water using a BP‐PSD on an Elekta Unity MR‐linac. Percentage depth doses (PDD) and profiles at different depths were measured at two movement velocities, 10 mm/s and 20 mm/s, for field sizes ranging from 10 × 10 cm 2 to 1 × 1 cm 2 . Gamma analysis was conducted to compare these measurements with those obtained during machine commissioning using a PTW Semiflex 3D ionization chamber (for PDD) and a PTW micro‐Diamond detector (for PDD and profiles). Gamma criteria of 2%/2 mm and 1%/1 mm dose difference/distance to agreement were studied, alongside field size, penumbra, and measurement time.

Results

All PDD and profile gamma passing rates were 100% at 2%/2 mm. At the stricter 1%/1 mm criteria, all PDD showed a passing rate above 96.97% for both velocities, with most of the profiles exceeding 95% at 10 mm/s and 90% at 20 mm/s. Gamma analysis results were superior for smaller fields (1 × 1 cm 2 and 2 × 2 cm 2 ) and generally better at 10 mm/s. On average, the penumbra measurements obtained with the PSD were greater than those achieved with the micro‐Diamond detector. Measurement times were found to be between 7 and 14 times shorter for PDD, and between 5 and 9 times shorter for profiles at speeds of 10 mm/s and 20 mm/s, respectively.

Conclusions

Ultra‐fast measurements using the Blue Physics PSD are suitable for acquiring dosimetric commissioning data with high accuracy and precision, and can be performed in a much shorter timeframe than with commonly used detectors.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Scintillation CountingPlasticsParticle AcceleratorsRadiotherapy Planning, Computer-AssistedRadiometryPhantoms, ImagingMagnetic Resonance ImagingHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

MR‐linacUltra‐fastcommissioningplastic scintillation detector

Substanzen

Plastics

Autor:innen und Institutionen

  • Department of Medical Physics and Radiation Protection, La Paz University Hospital, Paseo de la Castellana 261, Madrid, Spain.
  • Blue Physics LLC, 19145 long lake ranch blvd, Lutz, Florida, USA.

ORCID-Profile: Carlos Ferrer

Zitieren

APA

Ferrer, C., Huertas, C., Feijoo, M., Cardin, A., & Sáez, M. (2026). Ultra‐fast dosimetric data collection with a commercial plastic scintillation detector in an MR‐linac. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70460

Vancouver

Ferrer C, Huertas C, Feijoo M, Cardin A, Sáez M. Ultra‐fast dosimetric data collection with a commercial plastic scintillation detector in an MR‐linac. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70460

BibTeX
@article{Ferrer2026,
  author = {Ferrer, Carlos and Huertas, Concepción and Feijoo, Marcos and Cardin, Alessandro and Sáez, Moisés},
  title = {Ultra‐fast dosimetric data collection with a commercial plastic scintillation detector in an MR‐linac},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70460}
}
Kostenlos starten →