Beam‐axis stability during gantry rotation measured with scintillation dosimetry

Shingo Ohira, Takeshi Ohta iD, Honoka Sugaya, Tenyoh Suzuki iD, Yuki Nozawa, Masanari Minamitani, Atsuto Katano, Takuya Hayashi iD, Hideomi Yamashita, Keiichi Nakagawa

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70594 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Shingo Ohira und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, Graduate School of Human Health Sciences, Tokyo Metropolitan University, Arakawa-ku, Tokyo, Japan. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiotherapy, Intensity-Modulated, Scintillation Counting, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Volumetric modulated arc therapy (VMAT) requires high mechanical precision due to continuous gantry rotation. This study aims to investigate beam‐axis stability during VMAT delivery and evaluate scintillation‐based imaging as a complementary quality assurance (QA) tool.

Methods

Beam‐axis deviations were measured using a scintillation plate dosimeter, acquiring images at 5 frames per second in axial and coronal planes. Nine conformal arc beams with a field size of 1 × 1 cm 2 were delivered to a phantom under three configurations of the electric portal imaging device (EPID) and x‐ray volume imaging (XVI) systems: middle, retracted, and extended. Short‐ and long‐term reproducibility was assessed through repeated measurements. Deviations between the radiation dose center and the beam axis were analyzed in X, Y, and Z directions, and three‐dimensional (3D) displacements were calculated.

Results

Short‐term reproducibility showed maximum standard deviations (SDs) of 0.05, 0.06, and 0.11 mm for 4 MV, 6 MV, and 6 FFF beams, respectively. Long‐term reproducibility exhibited maximum SDs of 0.15, 0.13, and 0.19 mm, respectively. Angle‐dependent variations were most prominent at gantry angles of 0° and ±180°, consistent with gantry sag under gravity. The 3D displacements between the radiation dose center and the beam axis remained below 1 mm across all energies: the average (maximum) displacement ranged from 0.47 mm (0.96 mm) to 0.52 mm (0.97 mm) for 4 MV, 0.41 mm (0.68 mm) to 0.44 mm (0.76 mm) for 6 MV, and 0.42 mm (0.88 mm) to 0.47 mm (0.95 mm) for 6FFF. No relevant differences were observed among the three EPID/XVI configurations, indicating that these devices have a minimal impact on gantry mechanical stability.

Conclusions

Scintillation‐based imaging enables real‐time, high‐resolution evaluation of beam‐axis stability during VMAT. Deviations were consistently within submillimeter accuracy, supporting its use as a complementary QA method to conventional tests.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiotherapy, Intensity-ModulatedScintillation CountingRadiotherapy Planning, Computer-AssistedRadiometryQuality Assurance, Health CareParticle AcceleratorsHumansRadiotherapy DosageRotationReproducibility of Results

Stichwörter der Autor:innen

beam‐axisscintillationstability

Autor:innen und Institutionen

  • Department of Radiological Sciences, Graduate School of Human Health Sciences, Tokyo Metropolitan University, Arakawa-ku, Tokyo, Japan.
  • Department of Comprehensive Radiation Oncology, Graduate School of Medicine, The University of Tokyo, Bunkyo-ku, Tokyo, Japan.
  • Department of Radiology, The University of Tokyo Hospital, Bunkyo-ku, Tokyo, Japan.

ORCID-Profile: Takeshi Ohta, Tenyoh Suzuki, Takuya Hayashi

Zitieren

APA

Ohira, S., Ohta, T., Sugaya, H., Suzuki, T., Nozawa, Y., Minamitani, M., Katano, A., Hayashi, T., Yamashita, H., & Nakagawa, K. (2026). Beam‐axis stability during gantry rotation measured with scintillation dosimetry. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70594

Vancouver

Ohira S, Ohta T, Sugaya H, Suzuki T, Nozawa Y, Minamitani M, et al. Beam‐axis stability during gantry rotation measured with scintillation dosimetry. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70594

BibTeX
@article{Ohira2026,
  author = {Ohira, Shingo and Ohta, Takeshi and Sugaya, Honoka and Suzuki, Tenyoh and Nozawa, Yuki and Minamitani, Masanari and Katano, Atsuto and Hayashi, Takuya and Yamashita, Hideomi and Nakagawa, Keiichi},
  title = {Beam‐axis stability during gantry rotation measured with scintillation dosimetry},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70594}
}
Kostenlos starten →