EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Linac‐ and CyberKnife‐based MRI‐only treatment planning of prostate SBRT using an optimized synthetic CT calibration curve

Jessica Scholey iD, Tomi Nano iD, Kamal Singhrao iD, Osama Mohamad, Lisa Singer, Peder Eric Zufall Larson, Martina Descovich

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14411 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Jessica Scholey und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of California San Francisco, San Francisco, California, USA. Hauptthemen laut MEDLINE: Prostatic Neoplasms, Radiotherapy Planning, Computer-Assisted, Radiosurgery, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

CT Hounsfield Units (HUs) are converted to electron density using a calibration curve obtained from physical measurements of an electron density phantom. HU values assigned to an MRI‐derived synthetic computed tomography (sCT) may present a different relationship with electron density compared to CT HU. Correct assignment of sCT HU values is critical for accurate dose calculation and delivery. The goals of this work were to develop a sCT calibration curve using patient data acquired on a clinically commissioned CT scanner and assess for CyberKnife‐ and volumetric modulated arc therapy (VMAT)‐based MR‐only treatment planning of prostate SBRT.

Methods

Same‐day CT and MRI simulation in the treatment position were performed on 10 patients treated with SBRT to the prostate. Dixon in‐phase and out‐of‐phase MRIs were acquired on a 3T scanner using a 3D T1‐weighted gradient‐echo sequence to generate sCTs using a commercial sCT algorithm. CT and sCT datasets were co‐registered and HU values compared using mean absolute error (MAE). An optimized HU‐to‐density calibration curve was created based on average HU values across an institutional patient database for each of the four sCT tissue types. Clinical CyberKnife and VMAT treatment plans were generated on each patient CT and recomputed onto corresponding sCTs. Dose distributions computed using CT and sCT were compared using gamma criteria and dose‐volume‐histograms.

Results

For the optimized calibration curve, HU values were −96, 37, 204, and 1170 and relative electron densities were 0.95, 1.04, 1.1, and 1.7 for adipose, soft tissue, inner bone, and outer bone, respectively. The proposed sCT protocol produced total MAE of 94 ± 20HU. Gamma values mean ± std (min‐max) were 98.9% ± 0.9% (97.1%–100%) and 97.7% ± 1.3% (95.3%–99.3%) for VMAT and CyberKnife plans, respectively.

Conclusion

MRI‐derived sCT using the proposed approach shows excellent dosimetric agreement with conventional CT simulation, demonstrating the feasibility of MRI‐derived sCT for prostate SBRT treatment planning.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Prostatic NeoplasmsRadiotherapy Planning, Computer-AssistedRadiosurgeryRadiotherapy, Intensity-ModulatedRadiotherapy DosageMagnetic Resonance ImagingTomography, X-Ray ComputedPhantoms, ImagingHumansMaleCalibrationParticle AcceleratorsImage Processing, Computer-AssistedAlgorithmsOrgans at Risk

Stichwörter der Autor:innen

magnetic resonance imagingprostate cancerradiotherapysynthetic CTtreatment planning

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of California San Francisco, San Francisco, California, USA.
  • Department of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, California, USA.

ORCID-Profile: Jessica Scholey, Tomi Nano, Kamal Singhrao

Zitieren

APA

Scholey, J., Nano, T., Singhrao, K., Mohamad, O., Singer, L., Larson, P. E. Z., & Descovich, M. (2024). Linac‐ and CyberKnife‐based MRI‐only treatment planning of prostate SBRT using an optimized synthetic CT calibration curve. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14411

Vancouver

Scholey J, Nano T, Singhrao K, Mohamad O, Singer L, Larson PEZ, et al. Linac‐ and CyberKnife‐based MRI‐only treatment planning of prostate SBRT using an optimized synthetic CT calibration curve. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14411

BibTeX
@article{Scholey2024,
  author = {Scholey, Jessica and Nano, Tomi and Singhrao, Kamal and Mohamad, Osama and Singer, Lisa and Larson, Peder Eric Zufall and Descovich, Martina},
  title = {Linac‐ and CyberKnife‐based MRI‐only treatment planning of prostate SBRT using an optimized synthetic CT calibration curve},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14411}
}