Validation of the iFalcon V100 Rebound Tonometer in Canine Eyes

Yunhao Su, Yingbo Liu, Fushan Shi, Ji Fu iD

Veterinary Ophthalmology · 2026-03 · 29 (2)

DOI: 10.1111/vop.70168 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Veterinary Ophthalmology, online erschienen am 08.03.2026. Verfasst von Yunhao Su und 3 weiteren Autor:innen. Beteiligte Einrichtung: The School of Mechanical Engineering, Zhejiang Sci-Tech University, Hangzhou, China. Hauptthemen laut MEDLINE: Intraocular Pressure, Tonometry, Ocular. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 13 Literaturangaben. Gefördert von: Zhejiang Sci-Tech University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To validate the accuracy of the iFalcon V100 rebound tonometer in measuring intraocular pressure (IOP) in enucleated normal canine eyes. Animals Studied Four freshly enucleated normal canine eyes tested within 12 h postmortem. Procedures A 1.5 mL water syringe was inserted into the vitreous chamber, and an implantable pressure sensor was inserted into the anterior chamber. IOP measurements were initiated at 5 mmHg and gradually increased to 50 mmHg by incrementally advancing the syringe plunger. Measurements were taken with the iFalcon V100 from the central cornea using the canine mode, and results were compared to those obtained from the implantable pressure sensor using linear regression analysis and Bland–Altman plots.

Results

The iFalcon V100 readings showed a strong linear correlation with the implantable sensor ( r 2 = 0.96). Bland–Altman analysis revealed a mean bias of +0.34 mmHg, with 95% limits of agreement ranging from −4.29 to +4.98 mmHg; the overestimation was slightly in the lower pressure range. Notably, multiple attempts were required to obtain the measurements at 5–10 mmHg with the iFalcon V100.

Conclusions

The iFalcon V100 tonometer provides accurate central corneal IOP measurements in normal canine eyes within the physiological range of 5–50 mmHg, showing good agreement with direct implantable pressure sensor readings. Although the validated 5–50 mmHg range is wider than normal physiological pressures and encompasses commonly encountered pathological levels, we did not evaluate performance at 70–80 mmHg levels occurring in acute glaucoma, which is a study limitation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intraocular PressureTonometry, OcularAnimalsDogsReproducibility of Results

Stichwörter der Autor:innen

canineimplantable pressure sensorintraocular pressuremanometrytonometry accuracy

Autor:innen und Institutionen

  • The School of Mechanical Engineering, Zhejiang Sci-Tech University, Hangzhou, China.
  • Zhejiang University Center for Veterinary Sciences, Zhejiang University, Hangzhou, China.

ORCID-Profile: Ji Fu

Förderung

  • Zhejiang Sci-Tech University

Zitieren

APA

Su, Y., Liu, Y., Shi, F., & Fu, J. (2026). Validation of the iFalcon V100 Rebound Tonometer in Canine Eyes. Veterinary Ophthalmology, 29(2). https://doi.org/10.1111/vop.70168

Vancouver

Su Y, Liu Y, Shi F, Fu J. Validation of the iFalcon V100 Rebound Tonometer in Canine Eyes. Veterinary Ophthalmology. 2026;29(2). doi:10.1111/vop.70168

BibTeX
@article{Su2026,
  author = {Su, Yunhao and Liu, Yingbo and Shi, Fushan and Fu, Ji},
  title = {Validation of the iFalcon V100 Rebound Tonometer in Canine Eyes},
  journal = {Veterinary Ophthalmology},
  year = {2026},
  volume = {29},
  number = {2},
  doi = {10.1111/vop.70168}
}
Kostenlos starten →