Validation of the Reichert® Tono‐Vera® Vet rebound tonometer in normal ex vivo canine eyes
Veterinary Ophthalmology · 2024
DOI: 10.1111/vop.13213 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Veterinary Ophthalmology, 2024. Verfasst von Lydia E Kapeller und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA. Hauptthemen laut MEDLINE: Tonometry, Ocular, Intraocular Pressure. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 8 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NEI NIH HHS, NIH HHS, National Eye Institute.
Abstract (Originalsprache)
Objective
To determine the accuracy of the Reichert® Tono‐Vera® Vet rebound tonometer for canine intraocular pressure (IOP) measurement. Animals Studied Five normal canine ex vivo globes. Procedures The anterior chambers of five freshly enucleated normal canine eyes were cannulated and connected to a reservoir of Plasma‐Lyte A and a manometer. Starting at a manometric IOP of 5 mmHg, the pressure was progressively increased to 80 mmHg by raising the reservoir. Triplicate IOP measurements were taken with the Tono‐Vera® Vet from the central cornea using the dog setting and compared to the manometric pressure by linear regression analysis and Bland–Altman plots.
Results
There was a strong positive linear regression trend when comparing central corneal Tono‐Vera® Vet IOPs to manometric pressures ( r 2 = .99) with solid agreement between the two methods. Compared to manometric IOPs, the Tono‐Vera® Vet underestimated IOPs at higher pressures ≥70 mmHg.
Conclusions
Measurement of IOPs from the central cornea with the Tono‐Vera® Vet provided accurate results over a large range in normal canine globes compared to direct manometry. The mild to moderate underestimation of IOPs at high pressures was not considered clinically relevant.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA.
- Center for Statistical Training and Consulting, Michigan State University, East Lansing, Michigan, USA.
ORCID-Profile: András M Komáromy
Förderung
- NEI NIH HHS
- NIH HHS
- National Eye Institute
- BrightFocus Foundation
Zitieren
APA
Kapeller, L. E., Buckman, P. N., Wang, S., & Komáromy, A. M. (2024). Validation of the Reichert® Tono‐Vera® Vet rebound tonometer in normal ex vivo canine eyes. Veterinary Ophthalmology. https://doi.org/10.1111/vop.13213
Vancouver
Kapeller LE, Buckman PN, Wang S, Komáromy AM. Validation of the Reichert® Tono‐Vera® Vet rebound tonometer in normal ex vivo canine eyes. Veterinary Ophthalmology. 2024. doi:10.1111/vop.13213
BibTeX
@article{Kapeller2024,
author = {Kapeller, Lydia E and Buckman, Phillip N and Wang, Sichao and Komáromy, András M},
title = {Validation of the Reichert® Tono‐Vera® Vet rebound tonometer in normal ex vivo canine eyes},
journal = {Veterinary Ophthalmology},
year = {2024},
doi = {10.1111/vop.13213}
}
EUVIMED