Veterinary Ophthalmology · 2026
DOI: 10.1111/vop.70002 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Veterinary Ophthalmology, 2026. Verfasst von Jiwon Lee und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Veterinary Clinical Sciences, College of Veterinary Medicine and Research Institute for Veterinary Science, Seoul National University, …. Hauptthemen laut MEDLINE: Tonometry, Ocular, Intraocular Pressure. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Objective
To evaluate the accuracy of recalibrated TONOVET Plus (rTVP) and the newly released TONOVET Pro (TVPR) in measuring intraocular pressure (IOP) in ex vivo canine eyes compared to manometer and the established TONOVET (TV). Procedure Sixteen enucleated canine eyes were cannulated to establish manometric IOPs ranging from 5 to 80 mmHg. IOP measurements were obtained using TV, rTVP, and TVPR.
Results
All tonometers showed a strong correlation with the manometer ( R 2 > 0.98). Measurements from rTVP and TVPR exhibited almost perfect agreement with the manometer (CCC = 0.9974 for both), whereas TV demonstrated lower agreement (CCC = 0.9874) associated with greater systematic bias. Specifically, TV showed a lower proportional bias (95% confidence interval [CI] of the slope = 0.987–1.006) but a larger systematic bias (intercept = −3.33 mmHg) compared to rTVP and TVPR (95% CI of the slope = 1.045–1.066 and 1.045–1.063; intercept = −2.532 and −2.633 mmHg, respectively). In accuracy analyses by range, rTVP and TVPR showed no significant difference from the manometer in the 30–70 mmHg and 35–80 mmHg IOP ranges, respectively.
Conclusions
Among the ICare rebound tonometers, rTVP and TVPR provided the most accurate IOP measurements in the ex vivo analysis and demonstrated particularly strong agreement in the high IOP range, suggesting their suitability for IOP measurement in glaucomatous patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Veterinary Clinical Sciences, College of Veterinary Medicine and Research Institute for Veterinary Science, Seoul National University, Seoul, Korea.
- Keyprime Research Co., Ltd, Cheongju-si, Chungcheongbuk-do, Korea.
- Laboratory of Veterinary Ophthalmology, College of Veterinary Medicine, Gyeongsang National University, Jinju, Gyeongsangnam-do, Korea.
- Biotoxtech Co., Ltd., Cheongju-si, Chungcheongbuk-do, Korea.
ORCID-Profile: Jiwon Lee, Wontae Kim, Jaeho Shim, Yujin Kim, Seonmi Kang, Kangmoon Seo
Zitieren
APA
Lee, J., Kim, W., Shim, J., Kim, Y., Kang, S., Yoo, Y. K., Lee, Y. H., & Seo, K. (2026). Validation of Recalibrated TONOVET Plus and TONOVET Pro in Canine Eyes Compared to Manometer and TONOVET. Veterinary Ophthalmology. https://doi.org/10.1111/vop.70002
Vancouver
Lee J, Kim W, Shim J, Kim Y, Kang S, Yoo YK, et al. Validation of Recalibrated TONOVET Plus and TONOVET Pro in Canine Eyes Compared to Manometer and TONOVET. Veterinary Ophthalmology. 2026. doi:10.1111/vop.70002
BibTeX
@article{Lee2026,
author = {Lee, Jiwon and Kim, Wontae and Shim, Jaeho and Kim, Yujin and Kang, Seonmi and Yoo, Young-Kyoung and Lee, Young-Ho and Seo, Kangmoon},
title = {Validation of Recalibrated TONOVET Plus and TONOVET Pro in Canine Eyes Compared to Manometer and TONOVET},
journal = {Veterinary Ophthalmology},
year = {2026},
doi = {10.1111/vop.70002}
}
EUVIMED