Custom‐Made Polyethylene Tono‐Pen Tip Cover for Intraocular Pressure Measurement: Evaluation in a Controlled Eye Model and Canine Eyes
Veterinary Ophthalmology · 2026
DOI: 10.1111/vop.70134 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Ophthalmology, 2026. Verfasst von Pukkapol Suvannachart und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Faculty of Medicine, Khon Kaen University, Khon Kaen, Thailand. Hauptthemen laut MEDLINE: Intraocular Pressure, Tonometry, Ocular. Die Arbeit nennt 28 Literaturangaben.
Abstract (Originalsprache)
Objective
To evaluate the repeatability, reproducibility, agreement, and safety of intraocular pressure (IOP) measurements using a Tono‐Pen with a polyethylene tip cover (PC) compared to the latex tip cover (LC). Animal Studied A controlled eye model and healthy canine eyes. Procedures Measurements were performed in two phases. In the eye model, 55 intra‐balloon pressure levels (10 to 30 mmHg) were tested. Two measurements for each tip cover were performed to assess repeatability and agreement between the LC and PC. In canine eyes, repeatability and reproducibility of the PC were assessed using measurements from two examiners (77 eyes), while agreement between PC and LC was assessed using measurements by one examiner (79 eyes). Statistical analyses included mean differences (MD) and 95% limits of agreement (LOA) using the Bland–Altman method, and intraclass correlation coefficients (ICC).
Results
In the eye model, MD, LOA, and ICC were 0.02, −4.40 to 4.44, and 0.79 for PC repeatability; −0.25, −4.55 to 4.05, and 0.78 for LC repeatability; 0.56, −5.23 to 6.36, and 0.63 for LC and PC agreement. In canine eyes, MD, LOA, and ICC were −0.08, −4.61 to 4.47, and 0.84 for PC repeatability; −0.14, −5.44 to 5.16, and 0.77 for PC reproducibility; 0.13, −3.87 to 4.13, and 0.84 for LC and PC agreement. No corneal complications were observed in any measurements.
Conclusions
Compared to the conventional LC, the custom‐made PC proved to be a safe and reliable alternative tip cover for IOP measurement, demonstrating good repeatability, reproducibility, and agreement.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Ophthalmology, Faculty of Medicine, Khon Kaen University, Khon Kaen, Thailand.
- KKU Glaucoma Center of Excellence, Department of Ophthalmology, Faculty of Medicine, Khon Kaen University, Khon Kaen, Thailand.
- Veterinary Teaching Hospital, Khon Kaen University, Khon Kaen, Thailand.
- Division of Surgery, Faculty of Veterinary Medicine, Khon Kaen University, Khon Kaen, Thailand.
ORCID-Profile: Pukkapol Suvannachart
Zitieren
APA
Suvannachart, P., Asawaphureekorn, S., Sriphon, P., Lertitthikul, N., Jitasombuti, P., & Seesupa, S. (2026). Custom‐Made Polyethylene Tono‐Pen Tip Cover for Intraocular Pressure Measurement: Evaluation in a Controlled Eye Model and Canine Eyes. Veterinary Ophthalmology. https://doi.org/10.1111/vop.70134
Vancouver
Suvannachart P, Asawaphureekorn S, Sriphon P, Lertitthikul N, Jitasombuti P, Seesupa S. Custom‐Made Polyethylene Tono‐Pen Tip Cover for Intraocular Pressure Measurement: Evaluation in a Controlled Eye Model and Canine Eyes. Veterinary Ophthalmology. 2026. doi:10.1111/vop.70134
BibTeX
@article{Suvannachart2026,
author = {Suvannachart, Pukkapol and Asawaphureekorn, Somkiat and Sriphon, Phornrak and Lertitthikul, Nitiwadee and Jitasombuti, Preenun and Seesupa, Suvaluk},
title = {Custom‐Made Polyethylene Tono‐Pen Tip Cover for Intraocular Pressure Measurement: Evaluation in a Controlled Eye Model and Canine Eyes},
journal = {Veterinary Ophthalmology},
year = {2026},
doi = {10.1111/vop.70134}
}
EUVIMED