Calibration of a novel rebound tonometer ( Tono‐Vera ® vet) in the porcine eye

Erinn P Mills iD, Chin-Chi Liu, Ugochi Emelogu, Renee T Carter, Pilar Camacho-Luna iD, Andrew Christopher Lewin iD

Veterinary Ophthalmology · 2023

DOI: 10.1111/vop.13067 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Veterinary Ophthalmology, 2023. Verfasst von Erinn P Mills und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA. Hauptthemen laut MEDLINE: Tonometry, Ocular, Intraocular Pressure. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 12 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Porcine models of ocular disease are becoming increasingly utilized. A recently commercialized ocular tonometer, the Reichert Tono‐Vera® Vet, has not been evaluated for use in pigs. The purpose of this study was to calibrate this device for use in porcine eyes and to determine which settings are most appropriate for use in pigs. Procedure The anterior chambers of five freshly enucleated normal porcine eyes were cannulated then connected to a reservoir of balanced salt solution and a physiologic monitor. Triplicate measurements were obtained with the four available settings: dog, cat, horse, and rabbit at intraocular pressures ranging from 5– to 80 mmHg. Bland–Altman analysis was utilized to determine bias and 95% limits of agreement for each setting.

Results

There was a strong positive linear regression trend for all settings (dog r 2 = 0.986, horse r 2 = 0.947, cat r 2 = 0.977, and rabbit r 2 = 0.982). The linear regression equations for the dog, horse, cat, and rabbit setting were y = 1.0168x – 2.6128, y = 0.8743x – 3.4959, y = 0.9394x – 7.3188, and y = 1.1082x – 3.4077. The average bias and 95% limits of agreement for dog, horse, cat, and rabbit settings were − 2.00, −8.32, −9.58, and 0.57 mmHg, and (−7.52, 3.53), (−19.00, 2.37), (−16.66, −2.50), and (−7.79, 8.93), in mmHg.

Conclusion

The Tono‐Vera® Vet dog setting was most accurate and precise setting compared to true intraocular pressures. This setting is likely to be appropriate for in vivo use in pigs, with the appropriate correction formula applied.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tonometry, OcularIntraocular PressureAnimalsDogsSwineRabbitsHorsesCalibrationAnterior ChamberReproducibility of Results

Stichwörter der Autor:innen

Tono-Vera® Vetcalibrationintraocular pressurepigporcinetonometry

Autor:innen und Institutionen

  • Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.

ORCID-Profile: Erinn P Mills, Pilar Camacho-Luna, Andrew Christopher Lewin

Zitieren

APA

Mills, E. P., Liu, C. C., Emelogu, U., Carter, R. T., Camacho-Luna, P., & Lewin, A. C. (2023). Calibration of a novel rebound tonometer ( Tono‐Vera ® vet) in the porcine eye. Veterinary Ophthalmology. https://doi.org/10.1111/vop.13067

Vancouver

Mills EP, Liu CC, Emelogu U, Carter RT, Camacho-Luna P, Lewin AC. Calibration of a novel rebound tonometer ( Tono‐Vera ® vet) in the porcine eye. Veterinary Ophthalmology. 2023. doi:10.1111/vop.13067

BibTeX
@article{Mills2023,
  author = {Mills, Erinn P and Liu, Chin-Chi and Emelogu, Ugochi and Carter, Renee T and Camacho-Luna, Pilar and Lewin, Andrew Christopher},
  title = {Calibration of a novel rebound tonometer ( Tono‐Vera ® vet) in the porcine eye},
  journal = {Veterinary Ophthalmology},
  year = {2023},
  doi = {10.1111/vop.13067}
}
Kostenlos starten →