Validation of the eyeTelemed IOPvet indentation tonometer for use in dogs
Veterinary Ophthalmology · 2024
DOI: 10.1111/vop.13215 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Veterinary Ophthalmology, 2024. Verfasst von Lydia E Kapeller und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA. Hauptthemen laut MEDLINE: Tonometry, Ocular, Intraocular Pressure, Dog Diseases. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NEI NIH HHS, BrightFocus Foundation, NIH HHS.
Abstract (Originalsprache)
Objective
To assess the accuracy of canine intraocular pressure (IOP) estimates from the eyeTelemed IOPvet indentation tonometer. Animals Studied Part 1 included 54 eyes from 28 Beagle dogs—23 ADAMTS10 ‐mutants with open‐angle glaucoma and 5 normals. Part 2 involved five normal canine ex vivo globes. Procedure Part 1 (in vivo) compared IOPvet estimates in normal and glaucomatous dogs to Reichert Tono‐Vera® Vet rebound tonometry. The three IOPvet estimates were green (normal; 30 mmHg). In Part 2 (ex vivo), the pressure inside freshly enucleated normal canine eyes was progressively increased from 5 to 80 mmHg and compared to IOPvet estimates. Descriptive statistics compared IOPvet estimates to rebound tonometry and direct manometry, with the threshold from normal to glaucoma set at 30 mmHg.
Results
In Part 1 (in vivo), normal pressures (≤30 mmHg) were mainly identified correctly as green or yellow—110 of 111 estimates, corresponding to a specificity of 99%. Only 16 of 125 affected estimates were correctly displayed in the >30‐mmHg range; the remaining 109 showed ≤30 mmHg, corresponding to a sensitivity of 13%. In Part 2 (ex vivo), all normal pressures were correctly estimated with green, but 64 of 88 manometric IOPs >30 mmHg were falsely estimated as 20–30 mmHg.
Conclusions
The IOPvet is inaccurate in estimating canine IOP with a low sensitivity at identifying dogs with IOP > 30 mmHg. Canine‐specific instrument revision is required to correctly identify elevated (yellow = 20–30 mmHg) and high (red >30 mmHg) IOPs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Small Animal Clinical Sciences, College of Veterinary Medicine, Michigan State University, East Lansing, Michigan, USA.
- Center for Statistical Training and Consulting, Michigan State University, East Lansing, Michigan, USA.
ORCID-Profile: András M Komáromy
Förderung
- NEI NIH HHS
- BrightFocus Foundation
- NIH HHS
- Office of Undergraduate Research, Michigan State University
- Department of Animal Science, Michigan State University
- College of Agriculture and Natural Resources, Michigan State University
- Department of Entomology, College of Agriculture and Natural Resources, Michigan State University
- National Eye Institute
Zitieren
APA
Kapeller, L. E., Cabble, A. G., Buckman, P. N., Harman, C. D., Jacobson, A. L., Lawrence, F. R., & Komáromy, A. M. (2024). Validation of the eyeTelemed IOPvet indentation tonometer for use in dogs. Veterinary Ophthalmology. https://doi.org/10.1111/vop.13215
Vancouver
Kapeller LE, Cabble AG, Buckman PN, Harman CD, Jacobson AL, Lawrence FR, et al. Validation of the eyeTelemed IOPvet indentation tonometer for use in dogs. Veterinary Ophthalmology. 2024. doi:10.1111/vop.13215
BibTeX
@article{Kapeller2024,
author = {Kapeller, Lydia E and Cabble, Ava G and Buckman, Phillip N and Harman, Christine D and Jacobson, Amanda L and Lawrence, Frank R and Komáromy, András M},
title = {Validation of the eyeTelemed IOPvet indentation tonometer for use in dogs},
journal = {Veterinary Ophthalmology},
year = {2024},
doi = {10.1111/vop.13215}
}
EUVIMED