Validation of a novel rebound tonometer ( Tono‐Vera ® Vet) in normal ex vivo rabbit eyes
Veterinary Ophthalmology · 2023
DOI: 10.1111/vop.13142 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Ophthalmology, 2023. Verfasst von Erinn P Mills und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA. Hauptthemen laut MEDLINE: Tonometry, Ocular, Intraocular Pressure. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 12 Literaturangaben.
Abstract (Originalsprache)
Objective
The purpose of this study was to validate the use of the Reichert Tono‐Vera® Vet tonometer rabbit setting in normal ex vivo rabbit eyes and to compare the rabbit setting to the dog, cat, and horse settings of this tonometer. Procedure Six freshly enucleated normal rabbit eyes were cannulated and connected to a fluid reservoir and physiologic monitor. Triplicate measurements were obtained with the four available settings: dog, cat, horse, and rabbit at various intraocular pressures (IOP) ranging from 5 to 80 mmHg. Bland–Altman analysis was utilized to determine bias and 95% limits of agreement for each setting.
Results
Linear regression equations for the dog, horse, cat, and rabbit settings were y = 0.8101x + 2.5058, y = 0.7594x − 3.4673, y = 0.6635x + 0.3021, and y = 0.8935x + 1.3295, respectively. All settings demonstrated strong positive linear trends (dog r 2 = 0.9644, horse r 2 = 0.9456, cat r 2 = 0.9309, and rabbit r 2 = 0.9558). Bland–Altman plots revealed that the average bias and 95% limits of agreement (mmHg) were −4.73, −12.65, −12.86 and −2.73 and (−15.31, 5.86), (−29.03, 3.74), (−25.67, −0.05), and (−12.21, 6.76) for the dog, horse, cat, and rabbit settings, respectively.
Conclusion
The Tono‐Vera® Vet rabbit setting provided the most accurate and precise measurements compared with the other settings, but slightly underestimated actual IOP, especially as IOP was increased. This tonometer, using the rabbit setting, is likely to be appropriate for the estimation of IOP in rabbits with the appropriate correction formula applied.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.
ORCID-Profile: Erinn P Mills, Pilar Camacho-Luna, Andrew C Lewin
Zitieren
APA
Mills, E. P., Liu, C. C., Emelogu, U., Carter, R. T., Camacho-Luna, P., & Lewin, A. C. (2023). Validation of a novel rebound tonometer ( Tono‐Vera ® Vet) in normal ex vivo rabbit eyes. Veterinary Ophthalmology. https://doi.org/10.1111/vop.13142
Vancouver
Mills EP, Liu CC, Emelogu U, Carter RT, Camacho-Luna P, Lewin AC. Validation of a novel rebound tonometer ( Tono‐Vera ® Vet) in normal ex vivo rabbit eyes. Veterinary Ophthalmology. 2023. doi:10.1111/vop.13142
BibTeX
@article{Mills2023,
author = {Mills, Erinn P and Liu, Chin-Chi and Emelogu, Ugochi and Carter, Renee T and Camacho-Luna, Pilar and Lewin, Andrew C},
title = {Validation of a novel rebound tonometer ( Tono‐Vera ® Vet) in normal ex vivo rabbit eyes},
journal = {Veterinary Ophthalmology},
year = {2023},
doi = {10.1111/vop.13142}
}
EUVIMED