Comparison of three methods of tonometry in horses presented for ophthalmic disease
Veterinary Ophthalmology · 2023
DOI: 10.1111/vop.13135 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Ophthalmology, 2023. Verfasst von Emily Latham und 2 weiteren Autor:innen. Beteiligte Einrichtung: New Bolton Center, University of Pennsylvania School of Veterinary Medicine, Kennett Square, Pennsylvania, USA. Hauptthemen laut MEDLINE: Eye Diseases, Horse Diseases. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 13 Literaturangaben.
Abstract (Originalsprache)
Objective
To compare the measurement of intraocular pressure in horses with clinical ocular disease using three tonometry devices. Animals Studied A total of fifty horses were presented to the New Bolton Center Ophthalmology Service. Procedure Intraocular pressure was taken on 50 client‐owned horses (100 eyes) using the TonoVet, TonoVet Plus and Tono‐Pen Avia tonometers. Clinical equine patients included were presented to New Bolton Center for ophthalmic examination. Auriculopalpebral blocks were performed but horses were unsedated.
Results
All three tonometers were found to have strong agreement among them. The strongest agreement was between the TonoVet and the TonoVet Plus with the TonoVet on average 1.330 mmHg less than the TonoVet Plus (standard deviation 4.388 mmHg). This was followed by the TonoVet and the Tono‐Pen Avia Vet with the TonoVet on average 2.531 mmHg greater than the Tono‐Pen Avia Vet (standard deviation 4.124 mmHg). The weakest agreement was between the TonoVet Plus and the Tono‐Pen Avia with the TonoVet Plus on average 3.854 mmHg greater than the Tono‐Pen Avia (standard deviation 4.724 mmHg).
Conclusions
All three tonometers showed strong agreement, however, the TonoVet and the TonoVet Plus carried the strongest agreement with the TonoVet Plus having slightly higher measurements overall compared with the TonoVet. Due to small variations between devices, it is recommended that the same device be used for serial measurements of intraocular pressure. However, all three devices are appropriate to use in horses presented for ophthalmic evaluation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- New Bolton Center, University of Pennsylvania School of Veterinary Medicine, Kennett Square, Pennsylvania, USA.
ORCID-Profile: Emily Latham, Nicole M Scherrer
Zitieren
APA
Latham, E., Scherrer, N. M., & Stefanovski, D. (2023). Comparison of three methods of tonometry in horses presented for ophthalmic disease. Veterinary Ophthalmology. https://doi.org/10.1111/vop.13135
Vancouver
Latham E, Scherrer NM, Stefanovski D. Comparison of three methods of tonometry in horses presented for ophthalmic disease. Veterinary Ophthalmology. 2023. doi:10.1111/vop.13135
BibTeX
@article{Latham2023,
author = {Latham, Emily and Scherrer, Nicole M and Stefanovski, Darko},
title = {Comparison of three methods of tonometry in horses presented for ophthalmic disease},
journal = {Veterinary Ophthalmology},
year = {2023},
doi = {10.1111/vop.13135}
}
EUVIMED