Determination of intraocular pressure and Schirmer tear test and the comparison between the applanation tonometer (Tono‐Pen AVIA®) and the rebound tonometer (TonoVet Plus®) in mini lionhead rabbits
Veterinary Ophthalmology · 2024
DOI: 10.1111/vop.13150 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Ophthalmology, 2024. Verfasst von Ismael V Oliveira und 7 weiteren Autor:innen. Beteiligte Einrichtung: Graduate Program in Animal Science, Agricultural Science Center, Federal University of Paraíba (UFPB), Areia, Brazil. Hauptthemen laut MEDLINE: Intraocular Pressure, Eye Diseases. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 46 Literaturangaben.
Abstract (Originalsprache)
Objective
To determine intraocular pressure (IOP) and tear production, as well as to compare the IOP obtained with the TonoVet Plus® (rebound) with the Tono‐Pen Avia® (applanation) tonometers. Animals Twenty‐five Mini Lionhead rabbits ( n = 50 eyes). Procedure Tear production was measured at 6:00 a.m. and 6:00 p.m. by using the STT. The IOP reading was performed with the rebound tonometer, followed by the applanation tonometer, at 6:00 a.m., 9:00 a.m., 12:00 p.m., 3:00 p.m., and 6:00 p.m. Regression analysis, analysis of variance ( anova ) and Bland–Altman statistics were used.
Results
Daily tear production was 10.25 ± 3.75 mm/min, with no differences among the moments evaluated. Average daily IOP was 17.7 ± 3.08 mmHg with the TonoVet Plus® and 11.5 ± 4.56 mmHg with the Tono‐Pen Avia®. IOP values were higher at the beginning and end of the day with both tonometers.
Conclusion
The IOP values are higher with the TonoVet Plus® tonometer. The reference values of IOP and tear production obtained in this work may support the diagnosis, treatment, and monitoring of ocular disorders in pet Mini Lionhead rabbits.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Graduate Program in Animal Science, Agricultural Science Center, Federal University of Paraíba (UFPB), Areia, Brazil.
- Department of Veterinary Sciences, Agricultural Science Center, Federal University of Paraíba (UFPB), Areia, Brazil.
- Department of Animal Science, Agricultural Science Center, Federal University of Paraíba (UFPB), Areia, Brazil.
- Department of Social Sciences, Agricultural Science Center, Federal University of Paraíba (UFPB), Areia, Brazil.
ORCID-Profile: Danila B Campos
Zitieren
APA
Oliveira, I. V., Oliveira, L. V. S. X., Santos, V. M. B., Souza, J. L. T., Costa, M. L. L., Borges, P. F., Campos, D. B., & Talieri, I. C. (2024). Determination of intraocular pressure and Schirmer tear test and the comparison between the applanation tonometer (Tono‐Pen AVIA®) and the rebound tonometer (TonoVet Plus®) in mini lionhead rabbits. Veterinary Ophthalmology. https://doi.org/10.1111/vop.13150
Vancouver
Oliveira IV, Oliveira LVSX, Santos VMB, Souza JLT, Costa MLL, Borges PF, et al. Determination of intraocular pressure and Schirmer tear test and the comparison between the applanation tonometer (Tono‐Pen AVIA®) and the rebound tonometer (TonoVet Plus®) in mini lionhead rabbits. Veterinary Ophthalmology. 2024. doi:10.1111/vop.13150
BibTeX
@article{Oliveira2024,
author = {Oliveira, Ismael V and Oliveira, Lídia V S X and Santos, Victória M B and Souza, João L T and Costa, Maria Lindomárcia L and Borges, Péricles F and Campos, Danila B and Talieri, Ivia C},
title = {Determination of intraocular pressure and Schirmer tear test and the comparison between the applanation tonometer (Tono‐Pen AVIA®) and the rebound tonometer (TonoVet Plus®) in mini lionhead rabbits},
journal = {Veterinary Ophthalmology},
year = {2024},
doi = {10.1111/vop.13150}
}
EUVIMED