Journal article in Journal of Cardiovascular Electrophysiology, 2023. Written by Arash Aryana and 5 co-authors. Institution: Cardiac Arrhythmia Service, Mercy General Hospital and Dignity Health Heart and Vascular Institute, Sacramento, California, USA. Major MEDLINE topics: Pulmonary Veins, Atrial Fibrillation, Catheter Ablation. Cited 15 times so far (Crossref); 14 references. Funded by: CRC EP, Inc.
Abstract (original language)
Introduction
Pulsed field ablation (PFA) is a nonthermal ablative strategy that achieves cell death via electroporation. Herein, we investigated the preclinical safety and efficacy of PFA using two novel 8‐French, 16‐electrode spiral PFA/mapping catheters (ElePulse; CRC EP, Inc.).
Methods
Bipolar PFA (>1.8 kV) was performed using 30 s, single‐shot, QRS‐gated applications. Altogether, 94 atrial structures were ablated in 23 swine, one canine, and one ovine, including right and left atria and atrial appendages, pulmonary veins, and superior and inferior (IVC) vena cavae. We also examined the impact of PFA on the phrenic nerve (14 swine) and on a deviated esophagus after delivery of PFA from inside the IVC (five swine).
Results
All applications were single‐shot without catheter repositioning. Minimal microbubbling was observed without significant skeletal muscle twitching/activation (mean acceleration: 0.05 m/s 2 ). There was a marked reduction in post‐PFA versus pre‐PFA atrial electrogram amplitude (0.17 ± 0.21 vs. 1.18 ± 1.08 mV; p < .0001). Lesion durability was demonstrated up to 3 months in all targeted tissues. Histologically, lesions were contiguous and transmural, except in the atrial appendage, and without any thermal effects. Magnetic resonance, gross, and histologic examinations of the brain, rete mirabile, and kidneys revealed no thromboembolism. No acute/long‐term phrenic nerve dysfunction was encountered. Although within 2 h of ablation, histologic examinations of the esophagus revealed acute PFA‐related changes in the muscular layer, these completely resolved by 21 ± 5 days.
Conclusion
A novel, single‐shot, spiral PFA system is capable of safely creating large, durable atrial lesions without significant adverse effects on the phrenic nerve or the esophagus.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Cardiac Arrhythmia Service, Mercy General Hospital and Dignity Health Heart and Vascular Institute, Sacramento, California, USA.
- COR Healthcare, Torrance, California, USA.
- CRC EP, Inc., Tustin, California, USA.
ORCID profiles: Arash Aryana
Funding
- CRC EP, Inc
Cite
APA
Aryana, A., Ji, S. Y., Hata, C., de la Rama, A., Nguyen, K., & Panescu, D. (2023). Preclinical evaluation of a novel single‐shot pulsed field ablation system for pulmonary vein and atrial ablation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16010
Vancouver
Aryana A, Ji SY, Hata C, de la Rama A, Nguyen K, Panescu D. Preclinical evaluation of a novel single‐shot pulsed field ablation system for pulmonary vein and atrial ablation. Journal of Cardiovascular Electrophysiology. 2023. doi:10.1111/jce.16010
BibTeX
@article{Aryana2023,
author = {Aryana, Arash and Ji, Sang Yong and Hata, Cary and de la Rama, Alan and Nguyen, Ken and Panescu, Dorin},
title = {Preclinical evaluation of a novel single‐shot pulsed field ablation system for pulmonary vein and atrial ablation},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2023},
doi = {10.1111/jce.16010}
}
EUVIMED