Safety and Chronic Atrial Lesion Formation With a Large-Tip, Contact-Force-Sensing Multipolar Pulsed Field Ablation Catheter: Effect of Application Number in a Porcine Beating Heart Model
Circulation: Arrhythmia and Electrophysiology · 2026
DOI: 10.1161/circep.125.014473 ↗Kurz gefasst
Fachartikel in der Zeitschrift Circulation: Arrhythmia and Electrophysiology, 2026. Verfasst von Luigi Di Biase und 23 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Montefiore Medical Center, Bronx, NY (L.D.B., J.M., F.Z., D.Y., A.L., X.Z.). Hauptthemen laut MEDLINE: Heart Atria, Catheter Ablation, Cardiac Catheters, Pulmonary Veins. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
BACKGROUND: Integration of catheter–tissue contact force with pulsed field ablation (PFA) dosing is essential for achieving safe and durable lesion formation with contact force sensing catheters in ventricular models. However, data on the use of these catheters for atrial ablation remain limited. This study evaluated the procedural safety and 30-day lesion characteristics of atrial PFA delivered with the OMNYPULSE ablation catheter in a porcine model. METHODS: Twelve pigs were randomized to receive either 6 (×6 or Group A) or 12 (×12 or group B) PFA applications per ablation. Following 3-dimensional electroanatomic mapping, PFA was delivered to the right superior and inferior pulmonary veins, cavotricuspid isthmus, right atrial posterior wall, left atrial roof, and mitral annulus. Procedural safety was assessed acutely, and lesion characteristics were evaluated on 30-day histology. Maximum transmurality extent (MTE) was defined as >80% of the lesion span exhibiting continuous endocardial-to-epicardial fibrosis within representative histological sections. RESULTS: Adequate catheter–tissue contact was achieved during PFA delivery (contact force, 28±15 g). No acute procedural complications were observed. At 30 days, PVs and right atrial sites (ie, cavotricuspid isthmus and right atrial posterior wall) demonstrated near-complete MTE regardless of PFA dosing. In contrast, MTE was significantly lower at the mitral annulus and LA roof compared with other sites (44±30% and 21±26%, respectively; P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Montefiore Medical Center, Bronx, NY (L.D.B., J.M., F.Z., D.Y., A.L., X.Z.).
- Electrophysiology and Cardiac Pacing Unit, Humanitas Mater Domini, Castellanza, Varese, Italy (J.M.).
- Monzino Hospital Milan, Milano, Italy (M.S.).
- Bayside Preclinical Services Inc, Dixon (T. Spangler).
- Loma Linda University Heart Institute (R.B.).
- University of California, San Diego (J.H.).
ORCID-Profile: Luigi Di Biase, Jacopo Marazzato, Marco Schiavone, Fengwei Zou, Taylor Spangler, Rahul Bhardwaj, Jonathan Hsu, Assaf Govari, Andreas Altman, Kevin Okarski, Paras Parikh, Hiroshi Nakagawa, William Maddox, Devi Nair, Sanghamitra Mohanty, Vincenzo Mirco La Fazia, Andrea Natale, Xiaodong Zhang
Zitieren
APA
Di Biase, L., Marazzato, J., Schiavone, M., Zou, F., Spangler, T., Bhardwaj, R., Hsu, J., Govari, A., Altman, A., Okarski, K., Sharma, T., Parikh, P., Beeckler, C., Troutman, T., Chen, Q., Nakagawa, H., Maddox, W., Nair, D., Mohanty, S., …, & Zhang, X. (2026). Safety and Chronic Atrial Lesion Formation With a Large-Tip, Contact-Force-Sensing Multipolar Pulsed Field Ablation Catheter: Effect of Application Number in a Porcine Beating Heart Model. Circulation: Arrhythmia and Electrophysiology. https://doi.org/10.1161/circep.125.014473
Vancouver
Di Biase L, Marazzato J, Schiavone M, Zou F, Spangler T, Bhardwaj R, et al. Safety and Chronic Atrial Lesion Formation With a Large-Tip, Contact-Force-Sensing Multipolar Pulsed Field Ablation Catheter: Effect of Application Number in a Porcine Beating Heart Model. Circulation: Arrhythmia and Electrophysiology. 2026. doi:10.1161/circep.125.014473
BibTeX
@article{DiBiase2026,
author = {Di Biase, Luigi and Marazzato, Jacopo and Schiavone, Marco and Zou, Fengwei and Spangler, Taylor and Bhardwaj, Rahul and Hsu, Jonathan and Govari, Assaf and Altman, Andreas and Okarski, Kevin and Sharma, Tushar and Parikh, Paras and Beeckler, Christopher and Troutman, Terrence and Chen, Qi and Nakagawa, Hiroshi and Maddox, William and Nair, Devi and Mohanty, Sanghamitra and La Fazia, Vincenzo Mirco and Ynoa, Domingo and Lin, Aung and Natale, Andrea and Zhang, Xiaodong},
title = {Safety and Chronic Atrial Lesion Formation With a Large-Tip, Contact-Force-Sensing Multipolar Pulsed Field Ablation Catheter: Effect of Application Number in a Porcine Beating Heart Model},
journal = {Circulation: Arrhythmia and Electrophysiology},
year = {2026},
doi = {10.1161/circep.125.014473}
}
EUVIMED