Preclinical evaluation of a novel single‐shot pulsed field ablation system for pulmonary vein and atrial ablation
Journal of Cardiovascular Electrophysiology · 2023
DOI: 10.1111/jce.16010 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2023. Verfasst von Arash Aryana und 5 weiteren Autor:innen. Beteiligte Einrichtung: Cardiac Arrhythmia Service, Mercy General Hospital and Dignity Health Heart and Vascular Institute, Sacramento, California, USA. Hauptthemen laut MEDLINE: Pulmonary Veins, Atrial Fibrillation, Catheter Ablation. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 14 Literaturangaben. Gefördert von: CRC EP, Inc.
Abstract (Originalsprache)
Introduction
Pulsed field ablation (PFA) is a nonthermal ablative strategy that achieves cell death via electroporation. Herein, we investigated the preclinical safety and efficacy of PFA using two novel 8‐French, 16‐electrode spiral PFA/mapping catheters (ElePulse; CRC EP, Inc.).
Methods
Bipolar PFA (>1.8 kV) was performed using 30 s, single‐shot, QRS‐gated applications. Altogether, 94 atrial structures were ablated in 23 swine, one canine, and one ovine, including right and left atria and atrial appendages, pulmonary veins, and superior and inferior (IVC) vena cavae. We also examined the impact of PFA on the phrenic nerve (14 swine) and on a deviated esophagus after delivery of PFA from inside the IVC (five swine).
Results
All applications were single‐shot without catheter repositioning. Minimal microbubbling was observed without significant skeletal muscle twitching/activation (mean acceleration: 0.05 m/s 2 ). There was a marked reduction in post‐PFA versus pre‐PFA atrial electrogram amplitude (0.17 ± 0.21 vs. 1.18 ± 1.08 mV; p < .0001). Lesion durability was demonstrated up to 3 months in all targeted tissues. Histologically, lesions were contiguous and transmural, except in the atrial appendage, and without any thermal effects. Magnetic resonance, gross, and histologic examinations of the brain, rete mirabile, and kidneys revealed no thromboembolism. No acute/long‐term phrenic nerve dysfunction was encountered. Although within 2 h of ablation, histologic examinations of the esophagus revealed acute PFA‐related changes in the muscular layer, these completely resolved by 21 ± 5 days.
Conclusion
A novel, single‐shot, spiral PFA system is capable of safely creating large, durable atrial lesions without significant adverse effects on the phrenic nerve or the esophagus.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiac Arrhythmia Service, Mercy General Hospital and Dignity Health Heart and Vascular Institute, Sacramento, California, USA.
- COR Healthcare, Torrance, California, USA.
- CRC EP, Inc., Tustin, California, USA.
ORCID-Profile: Arash Aryana
Förderung
- CRC EP, Inc
Zitieren
APA
Aryana, A., Ji, S. Y., Hata, C., de la Rama, A., Nguyen, K., & Panescu, D. (2023). Preclinical evaluation of a novel single‐shot pulsed field ablation system for pulmonary vein and atrial ablation. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16010
Vancouver
Aryana A, Ji SY, Hata C, de la Rama A, Nguyen K, Panescu D. Preclinical evaluation of a novel single‐shot pulsed field ablation system for pulmonary vein and atrial ablation. Journal of Cardiovascular Electrophysiology. 2023. doi:10.1111/jce.16010
BibTeX
@article{Aryana2023,
author = {Aryana, Arash and Ji, Sang Yong and Hata, Cary and de la Rama, Alan and Nguyen, Ken and Panescu, Dorin},
title = {Preclinical evaluation of a novel single‐shot pulsed field ablation system for pulmonary vein and atrial ablation},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2023},
doi = {10.1111/jce.16010}
}
EUVIMED