Short implants: Influence of different diameters, torques, and transmucosal heights of straight prosthetic abutments on screw preload loss after cycling

Karla Regina Nogueira da Silva iD, Lucas Novaes Teixeira iD, Marcelo Lucchesi Teixeira iD, Guilherme Bovi Ambrosano iD, Elizabeth Ferreira Martinez iD

European Journal of Oral Sciences · 2023

DOI: 10.1111/eos.12937 ↗

At a glance

Journal article in European Journal of Oral Sciences, 2023. Written by Karla Regina Nogueira da Silva and 4 co-authors. Institution: Division of Oral Implantology, Faculdade São Leopoldo Mandic, Campinas, São Paulo, Brazil. Major MEDLINE topics: Dental Implant-Abutment Design, Dental Implants. 54 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

This study evaluated the influence of different implant diameters, insertion torques, and transmucosal heights on the loosening of abutments installed on short implants, after mechanical cycling. The Morse taper connection implants ( n = 96) tested were 5 mm high, divided according to the platform diameter: 4 or 6 mm. A universal abutment was coupled to each implant (with different transmucosal heights: 1 or 5 mm). The sets were subdivided into 20‐ and 32‐Ncm torque. After the cycle fatigue test, the detorque values were measured with a digital torque indicator. After mechanical cycling, the mean detorque values obtained for the abutment with 20‐Ncm insertion torque were lower than for implants with 32‐Ncm insertion torque, regardless of the platform diameter or transmucosal height. In the 20‐Ncm torque group, there was no statistically significant difference in the detorque values between platform diameters or transmucosal heights. Otherwise, for 32‐Ncm sets, a smaller platform diameter (4 mm), and a longer transmucosal height (5 mm) showed the lowest detorque values. In conclusion, implants placed with 32‐Ncm insertion torque and abutments with 1 mm transmucosal height and a 6 mm implant diameter demonstrated the highest detorque values.

Subjects

MeSH terms (MEDLINE)

Dental Implant-Abutment DesignDental ImplantsTorqueDental AbutmentsMaterials TestingStress, MechanicalBone ScrewsDental Stress Analysis

Author keywords

dental implantdental implant-abutment designloading

Substances

Dental Implants

Authors and institutions

  • Division of Oral Implantology, Faculdade São Leopoldo Mandic, Campinas, São Paulo, Brazil.
  • Division of Cell Biology, Faculdade São Leopoldo Mandic, Campinas, São Paulo, Brazil.
  • Division of Prosthodontics, Faculdade São Leopoldo Mandic, Campinas, São Paulo, Brazil.
  • Departament of Genetics, Laboratory of Statistical Genetics, Luiz de Queiroz College of Agriculture (ESALQ), University of São Paulo, São Paulo, Brazil.

ORCID profiles: Karla Regina Nogueira da Silva, Lucas Novaes Teixeira, Marcelo Lucchesi Teixeira, Guilherme Bovi Ambrosano, Elizabeth Ferreira Martinez

Cite

APA

da Silva, K. R. N., Teixeira, L. N., Teixeira, M. L., Ambrosano, G. B., & Martinez, E. F. (2023). Short implants: Influence of different diameters, torques, and transmucosal heights of straight prosthetic abutments on screw preload loss after cycling. European Journal of Oral Sciences. https://doi.org/10.1111/eos.12937

Vancouver

da Silva KRN, Teixeira LN, Teixeira ML, Ambrosano GB, Martinez EF. Short implants: Influence of different diameters, torques, and transmucosal heights of straight prosthetic abutments on screw preload loss after cycling. European Journal of Oral Sciences. 2023. doi:10.1111/eos.12937

BibTeX
@article{daSilva2023,
  author = {da Silva, Karla Regina Nogueira and Teixeira, Lucas Novaes and Teixeira, Marcelo Lucchesi and Ambrosano, Guilherme Bovi and Martinez, Elizabeth Ferreira},
  title = {Short implants: Influence of different diameters, torques, and transmucosal heights of straight prosthetic abutments on screw preload loss after cycling},
  journal = {European Journal of Oral Sciences},
  year = {2023},
  doi = {10.1111/eos.12937}
}