Fachartikel in der Zeitschrift Clinical Oral Implants Research, 2023. Verfasst von Gülce Çakmak und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland. Hauptthemen laut MEDLINE: Dental Cements, Dental Implants. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Buser Implant Foundation.
Abstract (Originalsprache)
Objectives
To investigate the effect of implant–abutment connection and screw channel angle on screw stability by comparing a newly introduced and an established connection, before and after cyclic loading. Materials and Methods Implants ( N = 44) with Torcfit (TF) or Crossfit (CF) connection were divided to be restored with a straight (CFS and TFS) or an angled screw access channel (CFA and TFA) titanium‐base abutment ( n = 11). CFA and TFA received screw‐retained crowns, whereas CFS and TFS received hybrid zirconia abutments and cement‐retained crowns. The initial torque value (ITV) of each complex (ITV I ) and removal torque value (RTV) after 24 h (RTV I ) were measured. Screws were replaced with new ones, ITVs were recorded again (ITV F ), and crowns were cyclically loaded (2.4 million cycles, 98 N) to measure RTVs again (RTV F ). Percentage torque loss was calculated. Data were analyzed ( α = 0.05).
Results
ITVs were similar among groups ( p ≥ .089). CF led to higher RTVs ( p ≤ .002), while CFS had higher RTV I than CFA ( p = .023). After 24 h, CFS had lower percentage torque loss than TF, while CFA had lower percentage torque loss than TFA ( p ≤ .011). After cyclic loading, CF led to lower percentage torque ( p < .001).
Conclusion
The implant–abutment connection affected the removal torque values. However, no screw loosening occurred during cyclic loading, which indicated a stable connection for all groups. Screw access channel angle did not affect screw stability after cyclic loading.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland.
- Department of Prosthodontics, Faculty of Dentistry, Necmettin Erbakan University, Konya, Turkey.
- Department of Prosthodontics, Faculty of Dentistry, Istinye University, Istanbul, Turkey.
- Ordu Oral and Dental Health Center, Ordu, Turkey.
- Division of Gerodontology and Removable Prosthodontics, University Clinics of Dental Medicine, University of Geneva, Geneva, Switzerland.
- Section of Digital Implant and Reconstructive Dentistry, Department of Reconstructive Dentistry and Gerodontology, University of Bern, Bern, Switzerland.
ORCID-Profile: Mustafa Borga Donmez, Martin Schimmel
Förderung
- Buser Implant Foundation
Zitieren
APA
Çakmak, G., Güven, M. E., Donmez, M. B., Kahveci, Ç., Schimmel, M., Abou-Ayash, S., & Yilmaz, B. (2023). Effect of internal connection type and screw channel angle on the screw stability of anterior implant‐supported zirconia crowns. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14165
Vancouver
Çakmak G, Güven ME, Donmez MB, Kahveci Ç, Schimmel M, Abou-Ayash S, et al. Effect of internal connection type and screw channel angle on the screw stability of anterior implant‐supported zirconia crowns. Clinical Oral Implants Research. 2023. doi:10.1111/clr.14165
BibTeX
@article{akmak2023,
author = {Çakmak, Gülce and Güven, Mehmet Esad and Donmez, Mustafa Borga and Kahveci, Çiğdem and Schimmel, Martin and Abou-Ayash, Samir and Yilmaz, Burak},
title = {Effect of internal connection type and screw channel angle on the screw stability of anterior implant‐supported zirconia crowns},
journal = {Clinical Oral Implants Research},
year = {2023},
doi = {10.1111/clr.14165}
}
EUVIMED