The Effects of Laboratory Contamination of Implant Abutment Screw and Connection on Reverse Torque Value – An In Vitro Study
Clinical and Experimental Dental Research · 2025-10 · 11 (5)
DOI: 10.1002/cre2.70222 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical and Experimental Dental Research, online erschienen am 16.09.2025. Verfasst von Hamed Bahrami Maleki und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics, School of Dentistry, Ilam University of Medical Sciences, Ilam, Iran. Hauptthemen laut MEDLINE: Dental Abutments, Equipment Contamination, Dental Implants, Bone Screws. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this study..
Abstract (Originalsprache)
Objectives
This in vitro experimental study aimed to evaluate the effects of laboratory contamination of implant abutment screws and connection surfaces on reverse torque values (RTVs), as an indicator of screw preload loss. Material and Methods Forty‐five Dentis One Q implants and 45 CCM UCLA abutments were randomly assigned into three groups ( n = 15 per group). Group 1 (control) involved uncontaminated abutments and screws with no restorations. Group 2 (screw contamination) used new abutments attached with screws contaminated by laboratory materials (porcelain powder, metal debris, and polishing paste). Group 3 (connection contamination) included screw‐retained restorations fabricated under contaminated conditions and attached using new screws. All samples were subjected to standardized torque (250 N·cm), thermocycling (1500 cycles between 5°C and 55°C), and subsequent RTV measurement. One‐way ANOVA and Tukey's post hoc tests were used for statistical analysis (α = 0.05).
Results
Mean RTVs (SD) were 218 (0.15) N·cm (control), 181 (0.14) N·cm (screw contamination), and 207 (0.11) N·cm (connection contamination). RTVs in the screw contamination group were significantly lower than both the control ( p < 0.001) and connection contamination groups ( p < 0.001). The difference between the control and connection contamination groups was not statistically significant ( p = 0.08).
Conclusions
Laboratory contamination of implant components can significantly reduce reverse torque values, particularly in cases of screw contamination, indicating an increased risk of screw loosening. Contamination control during prosthetic procedures is essential to maintaining implant stability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Prosthodontics, School of Dentistry, Ilam University of Medical Sciences, Ilam, Iran.
- General Dentist, Shiraz, Iran.
- Department of Prosthodontics, School of Dentistry, Kerman University of Medical Sciences, Kerman, Iran.
ORCID-Profile: Mona Bazooband
Förderung
- The authors received no specific funding for this study.
Zitieren
APA
Bahrami Maleki, H., Bazooband, M., & Amini, P. (2025). The Effects of Laboratory Contamination of Implant Abutment Screw and Connection on Reverse Torque Value – An In Vitro Study. Clinical and Experimental Dental Research, 11(5). https://doi.org/10.1002/cre2.70222
Vancouver
Bahrami Maleki H, Bazooband M, Amini P. The Effects of Laboratory Contamination of Implant Abutment Screw and Connection on Reverse Torque Value – An In Vitro Study. Clinical and Experimental Dental Research. 2025;11(5). doi:10.1002/cre2.70222
BibTeX
@article{BahramiMaleki2025,
author = {Bahrami Maleki, Hamed and Bazooband, Mona and Amini, Parviz},
title = {The Effects of Laboratory Contamination of Implant Abutment Screw and Connection on Reverse Torque Value – An In Vitro Study},
journal = {Clinical and Experimental Dental Research},
year = {2025},
volume = {11},
number = {5},
doi = {10.1002/cre2.70222}
}
EUVIMED