Improved organ sparing using auto‐planned Stanford volumetric modulated arc therapy for total body irradiation technique
Pediatric Blood & Cancer · 2023-10 · 70 (10)
DOI: 10.1002/pbc.30589 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pediatric Blood & Cancer, online erschienen am 24.07.2023. Verfasst von Nicholas Ngo und 6 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Oncology Department, Stanford University Cancer Center, Palo Alto, California, USA. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 19 Literaturangaben.
Abstract (Originalsprache)
Purpose/objectives To evaluate dosimetric differences between auto‐planned volumetric modulated arc therapy (VMAT) total body irradiation (TBI) technique and two‐dimensional radiotherapy using anterior‐posterial/posterio‐anterial beams (2D AP/PA) TBI technique.
Methods
Ten pediatric patients treated with VMAT‐TBI on Varian c‐arm linac were included in this study. VMAT‐TBI plans were generated using our in‐house developed and publicly shared auto‐planning scripts. For each VMAT‐TBI plan, a 2D AP/PA plan was created replicating the institution's clinical setup with the patient positioned at extended source to skin distance (SSD) with a compensator to account for differences in patient thickness, 50% transmission daily lung blocks, and electron chest wall boosts prescribed to 50% of the photon prescription. Clinically relevant metrics were analyzed and compared between the VMAT and 2D plans.
Results
All VMAT‐TBI plans achieved planned target volume (PTV) D90% ≥ 100% of prescription. VMAT‐TBI PTV D90% significantly increased (7.1% ± 2.9%, p < .001) compared to the 2D technique, whereas no differences were observed in global D max ( p < .2) and PTV V110% ( p < .4). Compared to the 2D plans, significant decreases in the D mean to the lungs (−25.6% ± 11.5%, p < .001) and lungs‐1 cm (−34.1% ± 10.1%, p < .001) were observed with the VMAT plans. The VMAT technique also enabled decrease of dose to other organs: kidneys D mean (−32.5% ± 5.0%, p < .001) and lenses D max (−5.3% ± 8.1%, p = .03); and in addition, for 2 Gy prescription: testes/ovaries D mean (−41.5% ± 11.5%, p < .001), brain D mean (−22.6% ± 5.4%, p = .002), and thyroid D mean (−18.2% ± 16.0%, p = .03).
Conclusions
Superior lung sparing with improved target coverage and similar global D max were observed with the VMAT plans as compared to 2D plans. In addition, VMAT‐TBI plans provided greater dose reductions in gonads, kidneys, brain, thyroid, and lenses.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Radiation Oncology Department, Stanford University Cancer Center, Palo Alto, California, USA.
ORCID-Profile: Susan M Hiniker, Nataliya Kovalchuk
Zitieren
APA
Ngo, N., Blomain, E. S., Simiele, E., Romero, I., Hoppe, R. T., Hiniker, S. M., & Kovalchuk, N. (2023). Improved organ sparing using auto‐planned Stanford volumetric modulated arc therapy for total body irradiation technique. Pediatric Blood & Cancer, 70(10). https://doi.org/10.1002/pbc.30589
Vancouver
Ngo N, Blomain ES, Simiele E, Romero I, Hoppe RT, Hiniker SM, et al. Improved organ sparing using auto‐planned Stanford volumetric modulated arc therapy for total body irradiation technique. Pediatric Blood & Cancer. 2023;70(10). doi:10.1002/pbc.30589
BibTeX
@article{Ngo2023,
author = {Ngo, Nicholas and Blomain, Erik S and Simiele, Eric and Romero, Ignacio and Hoppe, Richard T and Hiniker, Susan M and Kovalchuk, Nataliya},
title = {Improved organ sparing using auto‐planned Stanford volumetric modulated arc therapy for total body irradiation technique},
journal = {Pediatric Blood & Cancer},
year = {2023},
volume = {70},
number = {10},
doi = {10.1002/pbc.30589}
}
EUVIMED