Development of an external system for monitoring the couch speed in radiotherapy using continuous bed movement
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14497 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Hidetoshi Shimizu und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Aichi Cancer Center Hospital, Nagoya, Aichi, Japan. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging, Radiotherapy Dosage, Radiotherapy, Intensity-Modulated. Die Arbeit nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: JSPS KAKENHI.
Abstract (Originalsprache)
Purpose
Total body irradiation before bone marrow transplantation for hematological malignancies using Radixact, a high‐precision radiotherapy machine, can potentially reduce side effects and the risk of secondary malignancies. However, stable control of couch speed is critical, and direct assessment methods outlined in quality assurance guidelines are lacking. This study aims to develop a real‐time couch speed verification system for the Radixact.
Methods
The developed system used a linear encoder to measure couch speed directly. Accuracy was verified via a linear stage, comparing measurements with a laser distance sensor. After placing a phantom simulating the human body on the Radixact couch, the couch speed was verified using predefined speed plans.
Results
Operating the linear stage at 0.1, 0.5, and 1.0 mm/s revealed that the maximum position error of the developed verification system compared to the laser distance sensor was nearly equivalent to the distance resolution of the system (0.05 mm/pulse), with negligible average speed error. When the Radixact couch operated at 0.1, 0.5, and 1.0 mm/s, the values obtained by the verification system agreed with the theoretical values within the sampling period (0.01 s) and distance resolution (0.05 mm).
Conclusion
The verification system developed provides real‐time monitoring of the speed of the Radixact table, ensuring treatment effectiveness and patient safety. It would guarantee the couch speed's soundness and contribute to the “visualization” of safety.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Aichi Cancer Center Hospital, Nagoya, Aichi, Japan.
- QUALITA Co., Ltd., Matsumoto, Nagano, Japan.
- Graduate School of Radiological Technology, Gunma Prefectural College of Health Sciences, Maebashi, Gunma, Japan.
ORCID-Profile: Hidetoshi Shimizu, Takahiro Aoyama
Förderung
- JSPS KAKENHI
Zitieren
APA
Shimizu, H., Nakamura, O., Sasaki, K., Aoyama, T., Kitagawa, T., & Kodaira, T. (2024). Development of an external system for monitoring the couch speed in radiotherapy using continuous bed movement. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14497
Vancouver
Shimizu H, Nakamura O, Sasaki K, Aoyama T, Kitagawa T, Kodaira T. Development of an external system for monitoring the couch speed in radiotherapy using continuous bed movement. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14497
BibTeX
@article{Shimizu2024,
author = {Shimizu, Hidetoshi and Nakamura, Osamu and Sasaki, Koji and Aoyama, Takahiro and Kitagawa, Tomoki and Kodaira, Takeshi},
title = {Development of an external system for monitoring the couch speed in radiotherapy using continuous bed movement},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14497}
}
EUVIMED