Feasibility of segmental total body irradiation (SegTBI) using a 1.5T MR‐linac

Matthew Manhin Cheung, Ashley Chi Kin Cheng, Louis Lee

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70192 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Matthew Manhin Cheung und 2 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Division, Department of Medical Innovation & Technology, CUHK Medical Centre, Hong Kong SAR, China. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Whole-Body Irradiation, Radiotherapy, Intensity-Modulated, Phantoms, Imaging. Die Arbeit nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To evaluate the technical feasibility of using the elekta unity magnetic resonance linac (MRL) as a backup for tomotherapy in total body irradiation (TBI) treatment.

Methods

A single pediatric patient's TBI treatment with a prescription of 12 Gy in six fractions was retrospectively re‐planned using a multiple‐isocentre approach with dose feathering on the MRL system. An additional plan was created and delivered to an anthropomorphic phantom containing OSLDs. The study investigated the MRL system's limitations and capabilities.

Results

The plan sum of the nine segments not only met the dosimetric criteria of planning targets but also demonstrated the MRL system's capabilities by keeping the mean lung dose below 8 Gy and the mean kidney dose below 10 Gy. The electron streaming effect was observed. Treatment plan verification using ArcCHECK measurements with a global 3%/2 mm gamma analysis had a pass rate greater than 95% for all segments. In 28 out of the 30 OSLDs in brain, bone, and soft tissues, the deviation of the measurement from the reported TPS dose is within ±5%. A much larger deviation was observed in the lung tissues. Segmental TBI using the MRL was a viable option for TBI treatment. Significance This study demonstrates the technical feasibility of MRL for TBI by offering dose modulation and imaging capabilities. The MRL can serve as a backup despite longer planning and treatment times. The potential for future workflow optimizations could enhance its practicality. This research improves flexibility in treatment planning and delivery for TBI patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedWhole-Body IrradiationRadiotherapy, Intensity-ModulatedPhantoms, ImagingParticle AcceleratorsMagnetic Resonance ImagingOrgans at RiskHumansFeasibility StudiesRadiotherapy DosageRetrospective StudiesChild

Stichwörter der Autor:innen

MR Linacdose featheringtotal body irradiationunity

Autor:innen und Institutionen

  • Medical Physics Division, Department of Medical Innovation & Technology, CUHK Medical Centre, Hong Kong SAR, China.
  • CUHK Medical Centre, Hong Kong SAR, China.
  • Department of Clinical Oncology, Faculty of Medicine, CUHK, Hong Kong SAR, China.
  • Department of Medicine and Therapeutics, Faculty of Medicine, CUHK, Hong Kong SAR, China.

Zitieren

APA

Cheung, M. M., Cheng, A. C. K., & Lee, L. (2025). Feasibility of segmental total body irradiation (SegTBI) using a 1.5T MR‐linac. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70192

Vancouver

Cheung MM, Cheng ACK, Lee L. Feasibility of segmental total body irradiation (SegTBI) using a 1.5T MR‐linac. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70192

BibTeX
@article{Cheung2025,
  author = {Cheung, Matthew Manhin and Cheng, Ashley Chi Kin and Lee, Louis},
  title = {Feasibility of segmental total body irradiation (SegTBI) using a 1.5T MR‐linac},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70192}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche