Validation of Acuros for total body irradiation at extended distance

Geoffrey Nelson iD, Vikren Sarkar iD, Martin Szegedi iD, Andrea Molineu, Arthur J Olch, Jeremy N Kunz, Hui Zhao, Y Jessica Huang, Susha Pillai, Prema Rassiah iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14468 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Geoffrey Nelson und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Utah, Salt Lake City, Utah, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Algorithms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Standardized and accurately reported doses are essential in conventional total body irradiation (TBI), especially lung doses. This study evaluates the accuracy of the Acuros algorithm in predicting doses for extended‐distance TBI.

Methods

Measurements and calculations were done with both 6 and 18 MV. Tissue Maximum Ratio (TMR), output and off axis ratios (OAR) were measured at 200 and 500 cm source to detector distance and compared to Acuros calculated values. Two end‐to‐end tests were carried out, one with an in‐house phantom (solid water and Styrofoam) with inserted ion chambers and the other was with the Imaging and Radiation Oncology Core (IROC) TBI anthropomorphic phantom equipped with TLDs. The end‐to‐end test was done for 6 and 18 MV both with and without lung blocks. The source to midplane distance for both phantoms were at 518 and 508 cm respectively. Lung blocks were placed at the phantom surface and a beam spoiler was positioned 30 cm from the surface of the phantoms as per our clinical set up.

Results

The agreement between measured and calculated TMR, output and off axis ratios for both 6 and 18 MV were within 2%. Ion chamber measurements in both the Styrofoam and solid water for both energies carried out with and without lung blocks were within 2% of calculated values. TLD measured doses for both 6 and 18 MV in the IROC phantom were within 5% of calculated doses which is within the uncertainty of the TLD measurement.

Conclusions

The results indicate that the clinical beam model for Acuros 16.1 commissioned at standard clinical distances is capable of calculating doses accurately at extended distances up to 500 cm.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiotherapy Planning, Computer-AssistedRadiotherapy DosageAlgorithmsWhole-Body IrradiationHumansOrgans at RiskMonte Carlo MethodLungRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

AcurosEclipseTBI dose validationextended distance

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of Utah, Salt Lake City, Utah, USA.
  • Imaging and Radiation Oncology Core, Houston QA Center, MD Anderson Cancer Center, Houston, Texas, USA.
  • Department of Radiation Oncology, University of Southern California and Children's Hospital of Los Angeles, Los Angeles, California, USA.
  • Advent Health, Central Florida, Florida, USA.

ORCID-Profile: Geoffrey Nelson, Vikren Sarkar, Martin Szegedi, Prema Rassiah

Zitieren

APA

Nelson, G., Sarkar, V., Szegedi, M., Molineu, A., Olch, A. J., Kunz, J. N., Zhao, H., Huang, Y. J., Pillai, S., & Rassiah, P. (2024). Validation of Acuros for total body irradiation at extended distance. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14468

Vancouver

Nelson G, Sarkar V, Szegedi M, Molineu A, Olch AJ, Kunz JN, et al. Validation of Acuros for total body irradiation at extended distance. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14468

BibTeX
@article{Nelson2024,
  author = {Nelson, Geoffrey and Sarkar, Vikren and Szegedi, Martin and Molineu, Andrea and Olch, Arthur J and Kunz, Jeremy N and Zhao, Hui and Huang, Y Jessica and Pillai, Susha and Rassiah, Prema},
  title = {Validation of Acuros for total body irradiation at extended distance},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14468}
}
Kostenlos starten →