Quantifying the Transmit‐Power Threshold for Reliable Passive Detuning of Inductively Coupled Wireless Resonators in MRI: A Phantom Study

Ming Lu iD, Xiaoyu Jiang iD, Yajun Ma iD, Yurui Qian, Congyu Liao iD, Yang Yang iD, John C Gore iD, Xinqiang Yan iD

NMR in Biomedicine · 2026

DOI: 10.1002/nbm.70408 ↗

Kurz gefasst

Fachartikel in der Zeitschrift NMR in Biomedicine, 2026. Verfasst von Ming Lu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, Tennessee, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Magnetic Resonance Imaging, Wireless Technology. Die Arbeit nennt 48 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS, Vanderbilt Innovation Catalyst Fund.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This work develops and validates a practical method for quantifying the transmit‐power threshold at which passive detuning of inductively coupled wireless resonators becomes reliable in MRI. All experiments were performed using phantoms. A near/far spatial signal ratio is computed from a series of gradient‐echo (GRE) images acquired at different nominal flip angles. This ratio reveals ineffective detuning as a characteristic L‐shaped curve whose inflection point marks the detuning threshold. The method was validated experimentally with wireless resonators incorporating different numbers of crossed diode pairs, at multiple static field strengths (3 and 7T), and with different scanner vendors. At 3T, the method produced consistent thresholds across Philips and GE scanners despite their different flip‐angle thresholds. The proposed method provides a practical, straightforward tool for acceptance testing and quality assurance of wireless resonator safety in both clinical and research MRI settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingMagnetic Resonance ImagingWireless TechnologyReproducibility of ResultsEquipment DesignRadio Waves

Stichwörter der Autor:innen

RF safetyinductively coupledpassive detuningspatial signal ratiowireless resonator

Autor:innen und Institutionen

  • Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
  • Department of Radiology, University of California, San Diego, California, USA.
  • Department of Radiology and Biomedical Imaging, University of California, San Francisco, California, USA.
  • Department of Electrical and Computer Engineering, Vanderbilt University, Nashville, Tennessee, USA.

ORCID-Profile: Ming Lu, Xiaoyu Jiang, Yajun Ma, Congyu Liao, Yang Yang, John C Gore, Xinqiang Yan

Förderung

  • NIBIB NIH HHS
  • NIH HHS
  • Vanderbilt Innovation Catalyst Fund
  • National Institutes of Health

Zitieren

APA

Lu, M., Jiang, X., Ma, Y., Qian, Y., Liao, C., Yang, Y., Gore, J. C., & Yan, X. (2026). Quantifying the Transmit‐Power Threshold for Reliable Passive Detuning of Inductively Coupled Wireless Resonators in MRI: A Phantom Study. NMR in Biomedicine. https://doi.org/10.1002/nbm.70408

Vancouver

Lu M, Jiang X, Ma Y, Qian Y, Liao C, Yang Y, et al. Quantifying the Transmit‐Power Threshold for Reliable Passive Detuning of Inductively Coupled Wireless Resonators in MRI: A Phantom Study. NMR in Biomedicine. 2026. doi:10.1002/nbm.70408

BibTeX
@article{Lu2026,
  author = {Lu, Ming and Jiang, Xiaoyu and Ma, Yajun and Qian, Yurui and Liao, Congyu and Yang, Yang and Gore, John C and Yan, Xinqiang},
  title = {Quantifying the Transmit‐Power Threshold for Reliable Passive Detuning of Inductively Coupled Wireless Resonators in MRI: A Phantom Study},
  journal = {NMR in Biomedicine},
  year = {2026},
  doi = {10.1002/nbm.70408}
}
Kostenlos starten →