EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A Fast B1+‐Driven RF Co‐Simulation Optimization Method for RF Coil Arrays

Mazin M Mustafa iD, Ethan Polcyn, Matt Waks iD, Alexander Bratch, Gregory J Metzger iD, Xiufeng Li iD, Yigitcan Eryaman iD

Magnetic Resonance in Medicine · 2027-01 · 97 (1) · 441-453

DOI: 10.1002/mrm.70603 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 29.09.2026. Verfasst von Mazin M Mustafa und 6 weiteren Autor:innen. Beteiligte Einrichtung: Center for Magnetic Resonance Research, University of Minnesota at Twin Cities, Minneapolis, Minnesota, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS, NINDS NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To generate accurate EM simulation models of RF coil arrays, improve coil safety validation procedures, and improve RF power management without sacrificing patient safety.

Methods

A fast, customized RF co‐simulation formulation was developed for pTx RF coils, enabling the use of experimentally measured maps to optimize the values of lumped elements, phase shifters, and attenuators in the EM simulation model, and demonstrate in a phantom. The proposed method was tested on an 8‐channel pTx knee coil at 7T and a 16‐channel pTx head coil at 10.5T. An RF co‐simulation optimization method that relies solely on matching the simulated and experimental ‐matrices was also implemented as a baseline. 50,000 random RF shims were used to evaluate the error for both co‐simulation optimization methods.

Results

The proposed ‐driven co‐simulation optimization method generated accurate maps when compared to experimentally measured maps. The proposed method reduced the average NRMSE of the EM modeling against the measurements from 48.0% to 30.8% for the 7T 8‐channel coil, and from 91.1% to 50.9% for the 10.5T 16‐channel coil when compared to conventional methods using 50,000 random RF shims.

Conclusion

The developed ‐driven RF co‐simulation optimization method in this work can be used to generate a more accurate EM simulation model of custom‐built arrays and potentially reduce the need for large safety margins, with improved RF power management.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingImage Processing, Computer-AssistedPhantoms, ImagingComputer SimulationRadio WavesEquipment DesignHumansAlgorithmsReproducibility of Results

Stichwörter der Autor:innen

Co‐simulationMRIRF coilRF safetyoptimization

Autor:innen und Institutionen

  • Center for Magnetic Resonance Research, University of Minnesota at Twin Cities, Minneapolis, Minnesota, USA.

ORCID-Profile: Mazin M Mustafa, Matt Waks, Gregory J Metzger, Xiufeng Li, Yigitcan Eryaman

Förderung

  • NIBIB NIH HHS
  • NIH HHS
  • NINDS NIH HHS
  • National Institutes of Health

Zitieren

APA

Mustafa, M. M., Polcyn, E., Waks, M., Bratch, A., Metzger, G. J., Li, X., & Eryaman, Y. (2027). A Fast B1+‐Driven RF Co‐Simulation Optimization Method for RF Coil Arrays. Magnetic Resonance in Medicine, 97(1), 441-453. https://doi.org/10.1002/mrm.70603

Vancouver

Mustafa MM, Polcyn E, Waks M, Bratch A, Metzger GJ, Li X, et al. A Fast B1+‐Driven RF Co‐Simulation Optimization Method for RF Coil Arrays. Magnetic Resonance in Medicine. 2027;97(1):441-453. doi:10.1002/mrm.70603

BibTeX
@article{Mustafa2027,
  author = {Mustafa, Mazin M and Polcyn, Ethan and Waks, Matt and Bratch, Alexander and Metzger, Gregory J and Li, Xiufeng and Eryaman, Yigitcan},
  title = {A Fast B1+‐Driven RF Co‐Simulation Optimization Method for RF Coil Arrays},
  journal = {Magnetic Resonance in Medicine},
  year = {2027},
  volume = {97},
  number = {1},
  pages = {441--453},
  doi = {10.1002/mrm.70603}
}