Intra‐scan RF power amplifier drift correction
Magnetic Resonance in Medicine · 2024
DOI: 10.1002/mrm.30078 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von Ali Aghaeifar und 6 weiteren Autor:innen. Beteiligte Einrichtung: High-Field Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: ERC Advanced Grant, Deutsche Forschungsgemeinschaft (DFG), H2020 European Research Council.
Abstract (Originalsprache)
Purpose
The drift in radiofrequency (RF) power amplifiers (RFPAs) is assessed and several contributing factors are investigated. Two approaches for prospective correction of drift are proposed and their effectiveness is evaluated.
Methods
RFPA drift assessment encompasses both intra‐pulse and inter‐pulse drift analyses. Scan protocols with varying flip angle (FA), RF length, and pulse repetition time (TR) are used to gauge the influence of these parameters on drift. Directional couplers (DICOs) monitor the forward waveforms of the RFPA outputs. DICOs data is stored for evaluation, allowing calculation of correction factors to adjust RFPAs' transmit voltage. Two correction methods, predictive and run‐time, are employed: predictive correction necessitates a calibration scan, while run‐time correction calculates factors during the ongoing scan.
Results
RFPA drift is indeed influenced by the RF duty‐cycle, and in the cases examined with a maximum duty‐cycle of 66%, the potential drift is approximately 41% or 15%, depending on the specific RFPA revision. Notably, in low transmit voltage scenarios, FA has minimal impact on RFPA drift. The application of predictive and run‐time drift correction techniques effectively reduces the average drift from 10.0% to less than 1%, resulting in enhanced MR signal stability.
Conclusion
Utilizing DICO recordings and implementing a feedback mechanism enable the prospective correction of RFPA drift. Having a calibration scan, predictive correction can be utilized with fewer complexity; for enhanced performance, a run‐time approach can be employed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- High-Field Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany.
- Department of Biomedical Magnetic Resonance, University of Tuebingen, Tuebingen, Germany.
- Center for MR Research, University Children's Hospital, Zurich, Switzerland.
- Imaging Centre of Excellence, University of Glasgow, Glasgow, UK.
- MR Physics, German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany.
- NeuroSpin, Paris-Saclay University, Gif-sur-yvette, France.
ORCID-Profile: Ali Aghaeifar, Dario Bosch, Rahel Heule, Sydney Williams, Philipp Ehses, Franck Mauconduit, Klaus Scheffler
Förderung
- ERC Advanced Grant
- Deutsche Forschungsgemeinschaft (DFG)
- H2020 European Research Council
Zitieren
APA
Aghaeifar, A., Bosch, D., Heule, R., Williams, S., Ehses, P., Mauconduit, F., & Scheffler, K. (2024). Intra‐scan RF power amplifier drift correction. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30078
Vancouver
Aghaeifar A, Bosch D, Heule R, Williams S, Ehses P, Mauconduit F, et al. Intra‐scan RF power amplifier drift correction. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30078
BibTeX
@article{Aghaeifar2024,
author = {Aghaeifar, Ali and Bosch, Dario and Heule, Rahel and Williams, Sydney and Ehses, Philipp and Mauconduit, Franck and Scheffler, Klaus},
title = {Intra‐scan RF power amplifier drift correction},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30078}
}
EUVIMED