Revisiting Inductively Coupled Wireless Coils in MRI : Mitigating Over‐Coupling With Preamplifiers
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70485 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Ming Lu und 2 weiteren Autor:innen. Beteiligte Einrichtung: Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, Tennessee, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Wireless Technology. Die Arbeit nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIBIB NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Purpose
Inductively coupled coils enhance local MRI sensitivity, yet strong coupling with nearby primary coils typically causes resonance splitting and impedance mismatch, which are traditionally considered detrimental. This work investigates why inductively coupled coils can still function effectively even in the presence of severe coupling and clarifies the role of modern receive preamplifiers in mitigating coupling effects.
Methods
Bench experiments were performed using primary coils (10 and 15 cm) and secondary inductively coupled coils (3–9 cm) tuned to the same Larmor frequency at 1.5, 3, and 7 T. Resonance characteristics and primary‐coil impedance variations were evaluated under open‐circuit, 50‐Ω, and low‐input‐impedance preamplifier terminations. MRI validation was conducted at 7 T without retuning the primary coil after introducing a closely positioned, inductively coupled coil, while intentionally varying preamplifier decoupling conditions.
Results
Open‐circuit and 50‐Ω terminations produced pronounced resonance splitting and significant impedance distortion. In contrast, low‐input‐impedance preamplifier termination preserved the inductively coupled coil resonance despite strong coupling. Although the primary‐coil impedance shifted substantially, it remained within acceptable noise‐figure contours, resulting in negligible SNR penalty. Degraded preamplifier decoupling led to a 21%–23% SNR reduction.
Conclusion
Modern preamplifiers fundamentally alter coupled‐coil behavior, enabling inductively coupled coils to operate near primary coils without significant SNR degradation and simplifying inductively coupled coil design.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Vanderbilt University Institute of Imaging Science, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
- Department of Biomedical Engineering, Vanderbilt University, Nashville, Tennessee, USA.
- Department of Radiology and Radiological Sciences, Vanderbilt University Medical Center, Nashville, Tennessee, USA.
- Department of Electrical and Computer Engineering, Vanderbilt University, Nashville, Tennessee, USA.
ORCID-Profile: Ming Lu, Xinqiang Yan
Förderung
- NIBIB NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Lu, M., Gore, J. C., & Yan, X. (2026). Revisiting Inductively Coupled Wireless Coils in MRI : Mitigating Over‐Coupling With Preamplifiers. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70485
Vancouver
Lu M, Gore JC, Yan X. Revisiting Inductively Coupled Wireless Coils in MRI : Mitigating Over‐Coupling With Preamplifiers. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70485
BibTeX
@article{Lu2026,
author = {Lu, Ming and Gore, John C and Yan, Xinqiang},
title = {Revisiting Inductively Coupled Wireless Coils in MRI : Mitigating Over‐Coupling With Preamplifiers},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70485}
}
EUVIMED