Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 02.10.2026. Verfasst von Oskar Björkqvist und 1 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Enhancement. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Swiss National Science Foundation.
Abstract (Originalsprache)
Purpose
Recent work has proposed self‐synchronization, self‐calibration, and means of sequence parsing to enhance the versatility of field cameras. The purpose of this paper is to expand this approach to more complex sequences and verify feasibility across distinct scanner platforms.
Methods
A field camera was formed from six field probes in a rigid‐body arrangement and deployed in two commercial MRI scanners. Epoch folding for synchronization and the formation of sequence diagrams were expanded to nested sequences. Autonomous camera operation was used to record advanced sequences involving gradient‐echo trains, spin‐echo trains, and diffusion‐tensor encoding.
Results
Autonomous camera operation was successful in both scanners, yielding sequence diagrams for all sequences explored.
Conclusion
Autonomy of field cameras permits use across different MRI systems and degrees of sequence complexity. Potential applications range from method development and validation to sequence harmonization, multi‐center studies, and quality assurance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute for Biomedical Engineering, ETH Zurich and University of Zurich, Zurich, Switzerland.
ORCID-Profile: Oskar Björkqvist, Klaas P Pruessmann
Förderung
- Swiss National Science Foundation
Zitieren
APA
Björkqvist, O., & Pruessmann, K. P. (2027). Cross‐Platform Demonstration of an Autonomous MRI Field Camera. Magnetic Resonance in Medicine, 97(1), 464-471. https://doi.org/10.1002/mrm.70601
Vancouver
Björkqvist O, Pruessmann KP. Cross‐Platform Demonstration of an Autonomous MRI Field Camera. Magnetic Resonance in Medicine. 2027;97(1):464-471. doi:10.1002/mrm.70601
BibTeX
@article{Bjrkqvist2027,
author = {Björkqvist, Oskar and Pruessmann, Klaas P},
title = {Cross‐Platform Demonstration of an Autonomous MRI Field Camera},
journal = {Magnetic Resonance in Medicine},
year = {2027},
volume = {97},
number = {1},
pages = {464--471},
doi = {10.1002/mrm.70601}
}
EUVIMED