Design of an 8‐Channel Transmit 32‐Channel Receive 11.7T Head Coil and Evaluation of SNR Gains
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70473 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Son Chu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Imaging Centre of Excellence, University of Glasgow, Glasgow, UK. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Transducers. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Agence Nationale de la Recherche, European Union's Horizon 2020 research and innovation programme, Horizon 2020 Framework Programme.
Abstract (Originalsprache)
Purpose
To design and develop an 8‐channel transmit 32‐channel receive 11.7T head coil for the strongest human MRI scanner to date, demonstrate safety, and evaluate signal‐to‐noise ratio (SNR) gains compared to 7T, the most used ultra‐high‐field (UHF) platform.
Methods
The transmit array design incorporated a folded‐end radiofrequency (RF) shield to mitigate wave propagation in the scanner bore. The receive array consisted of 32‐overlapped loops constructed on a helmet size‐matched with the industry standard 7T head coil. The transmit array was characterized using B 1 + mapping and MR thermometry, whereas the receive array performance was evaluated by measuring SNR at 7T and 11.7T using a lightbulb‐shaped phantom filled with a tissue equivalent solution.
Results
Loss due to wave propagation was 4.5% of the input power for the chosen RF shield. The mean magnitude of the complex correlations between the simulated and measured single channel B 1 + maps was 94%, which highlights minimal interaction between the transmit and receive arrays. An SNR gain of three‐fold was achieved in the phantom centre. In the periphery, representing the cortex, the SNR gain was 2.65‐fold, and the average gain over the whole volume was a factor of 2.69. The g‐factors of the 11.7T array were lower than the reference 7T coil due to the distinct coil sensitivities at higher frequency.
Conclusion
We report substantial SNR gains over the whole volume at 11.7T using a fully overlapped 32‐channel loop array. The central SNR gain of B 0 2.14 agrees with the theoretical prediction, while we also gain supra‐linearly in the periphery.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Imaging Centre of Excellence, University of Glasgow, Glasgow, UK.
- MR CoilTech Limited, Glasgow, UK.
- University of Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif sur Yvette, France.
- MR Physics, German Center for Neurodegenerative Diseases (DZNE), Bonn, Germany.
ORCID-Profile: Son Chu, Caroline Le Ster, Alexis Amadon, Vincent Gras, Paul-François Gapais, Rüdiger Stirnberg, Nicolas Boulant, Shajan Gunamony
Förderung
- Agence Nationale de la Recherche
- European Union's Horizon 2020 research and innovation programme
- Horizon 2020 Framework Programme
Zitieren
APA
Chu, S., Issac, J. P., Le Ster, C., Amadon, A., Gras, V., Gapais, P. F., Stirnberg, R., Boulant, N., & Gunamony, S. (2026). Design of an 8‐Channel Transmit 32‐Channel Receive 11.7T Head Coil and Evaluation of SNR Gains. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70473
Vancouver
Chu S, Issac JP, Le Ster C, Amadon A, Gras V, Gapais PF, et al. Design of an 8‐Channel Transmit 32‐Channel Receive 11.7T Head Coil and Evaluation of SNR Gains. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70473
BibTeX
@article{Chu2026,
author = {Chu, Son and Issac, Jeslin Parackal and Le Ster, Caroline and Amadon, Alexis and Gras, Vincent and Gapais, Paul-François and Stirnberg, Rüdiger and Boulant, Nicolas and Gunamony, Shajan},
title = {Design of an 8‐Channel Transmit 32‐Channel Receive 11.7T Head Coil and Evaluation of SNR Gains},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70473}
}
EUVIMED