Coaxial Dipole Array With Switching Transmit Sensitivities for Ultrahigh Field MRI
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70243 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Dario Bosch und 5 weiteren Autor:innen. Beteiligte Einrichtung: Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Enhancement, Transducers, Signal Processing, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft, European Research Council.
Abstract (Originalsprache)
Purpose
To investigate dipole antennas with electronically switchable transmit field patterns to improve flip angle homogeneity in ultra‐high field MRI.
Methods
Reconfigurable dipole elements that could produce two distinct electronically switchable field profiles were conceptualized and constructed. Eight such elements were combined into an array. Alteration of the field profiles was accomplished by modulating the currents along the dipoles using a combination of PIN diodes and lumped inductors. The behavior of these reconfigurable elements was studied in numerical electromagnetic simulations and 9.4T MRI measurements, investigating rapid switching of transmit sensitivities during excitation pulses in both single‐channel and pTx mode operation.
Results
For the simulated dipole elements, modulating the current densities along the dipole's axis causes a 30% change of the field between superior and inferior regions of the brain. When rapidly switched during excitation pulses, this degree of freedom can improve flip angle homogeneity, e.g., by a factor of 2.2 for a two kT points pTx pulse. For the constructed prototype array, the switching effect was observable but weaker, causing 10% superior‐inferior variation.
Conclusions
The proposed coaxial dipole array with switchable transmit sensitivities offers a novel degree of freedom for designing excitation pulses. The approach has the potential to improve flip angle homogeneity without necessitating an expensive increase in the number of independent transmit channels.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
- Biomedical Magnetic Resonance, University of Tübingen, Tübingen, Germany.
- MRI Core Facility of the Medical Faculty, University of Tübingen, Tübingen, Germany.
- Institute of Biomedical Imaging, Graz University of Technology, Graz, Austria.
- Institute of Neuroradiology, University Hospital Erlangen, Erlangen, Germany.
ORCID-Profile: Dario Bosch, Georgiy A Solomakha, Felix Glang, Nikolai Avdievich, Klaus Scheffler
Förderung
- Deutsche Forschungsgemeinschaft
- European Research Council
Zitieren
APA
Bosch, D., Solomakha, G. A., Glang, F., Freudensprung, M., Avdievich, N., & Scheffler, K. (2026). Coaxial Dipole Array With Switching Transmit Sensitivities for Ultrahigh Field MRI. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70243
Vancouver
Bosch D, Solomakha GA, Glang F, Freudensprung M, Avdievich N, Scheffler K. Coaxial Dipole Array With Switching Transmit Sensitivities for Ultrahigh Field MRI. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70243
BibTeX
@article{Bosch2026,
author = {Bosch, Dario and Solomakha, Georgiy A and Glang, Felix and Freudensprung, Martin and Avdievich, Nikolai and Scheffler, Klaus},
title = {Coaxial Dipole Array With Switching Transmit Sensitivities for Ultrahigh Field MRI},
journal = {Magnetic Resonance in Medicine},
year = {2026},
doi = {10.1002/mrm.70243}
}
EUVIMED