Improved Myocardial Sodium Quantification at 7 T Using Interleaved 23 Na / 1 H pTx MRI With Motion and Anatomy‐Based B 1 Correction

Laurent Ruck iD, Nico Egger iD, Benedikt Zobler, Judith Schirmer, Sophia Nagelstraßer, Andreas Bitz, Tanja Platt iD, Simon Konstandin iD, Christoph Kopp, Michael Uder, Armin Michael Nagel iD

Magnetic Resonance in Medicine · 2026-07 · 96 (1) · 173-190

DOI: 10.1002/mrm.70342 ↗

At a glance

Journal article in Magnetic Resonance in Medicine, published online on 2026-03-16. Written by Laurent Ruck and 10 co-authors. Institution: Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany. Major MEDLINE topics: Sodium, Heart, Magnetic Resonance Imaging, Myocardium. Cited 1 times so far (Crossref); 48 references. The full text is freely available (open access, CC BY 4.0). Funded by: Deutsche Forschungsgemeinschaft.

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Purpose

To improve accuracy and repeatability of myocardial 23 Na quantification in cardiac MRI at 7 T by combining retrospective respiratory and cardiac motion correction with a novel anatomy‐based B 1 bias field correction.

Methods

In this study, a dual‐nuclear interleaved 23 Na/ 1 H MRI sequence at 7 T was applied. Here, 1 H MR data enabled automated segmentation of myocardium and blood pool via nnUNet and facilitated respiratory and cardiac motion correction for both contrasts using 1 H navigators. For 23 Na MRI, a novel anatomy‐based B 1 bias field correction was developed, estimating the transmit and receive (B 1 ) field bias from 1 H‐derived segmentations and synthetic 23 Na images. All correction methods were validated through realistic simulations and in vivo studies. Repeatability was assessed in 10 healthy subjects, and the proposed B 1 correction was compared to phantom‐based methods.

Results

Respiratory and cardiac motion correction reduced quantification errors and improved SNR compared to conventional gating. The anatomy‐based B 1 correction effectively mitigated errors caused by B 1 inhomogeneities and outperformed phantom‐based methods in terms of repeatability. The measured apparent tissue sodium concentration of the myocardium after all corrections (49.5 4.7 mM) was consistent with literature values. Combined motion and B 1 corrections improved repeatability, reducing the coefficient of repeatability from 4.9 mM (11.2%) without corrections to 2.0 mM (4.0%) after corrections.

Conclusion

The combination of retrospective motion correction and anatomy‐based B 1 bias field correction enables repeatable myocardial sodium quantification at 7 T. The interleaved acquisition facilitates robust segmentation and efficient use of 1 H MRI for motion and B 1 correction, providing a scalable framework for future clinical research studies.

Subjects

MeSH terms (MEDLINE)

SodiumHeartMagnetic Resonance ImagingMyocardiumImage Processing, Computer-AssistedHumansReproducibility of ResultsMotionAdultPhantoms, ImagingMaleFemaleAlgorithmsSodium IsotopesSignal-To-Noise RatioRespiration

Author keywords

cardiac sodium (23Na) MRIinterleaved dual‐nuclear MRIrespiratory and cardiac motion correctiontransmit and receive (B1) correctionultrahigh field strength (7 T)

Substances

SodiumSodium Isotopes

Authors and institutions

  • Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Department of Nephrology and Hypertension, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Electrical Engineering and Information Technology, University of Applied Sciences - FH Aachen, Aachen, Germany.
  • Medical Physics in Radiology, German Cancer Research Center (DKFZ) Heidelberg, Heidelberg, Germany.
  • Fraunhofer Institute for Digital Medicine MEVIS, Bremen, Germany.

ORCID profiles: Laurent Ruck, Nico Egger, Tanja Platt, Simon Konstandin, Armin Michael Nagel

Funding

  • Deutsche Forschungsgemeinschaft

Cite

APA

Ruck, L., Egger, N., Zobler, B., Schirmer, J., Nagelstraßer, S., Bitz, A., Platt, T., Konstandin, S., Kopp, C., Uder, M., & Nagel, A. M. (2026). Improved Myocardial Sodium Quantification at 7 T Using Interleaved 23 Na / 1 H pTx MRI With Motion and Anatomy‐Based B 1 Correction. Magnetic Resonance in Medicine, 96(1), 173-190. https://doi.org/10.1002/mrm.70342

Vancouver

Ruck L, Egger N, Zobler B, Schirmer J, Nagelstraßer S, Bitz A, et al. Improved Myocardial Sodium Quantification at 7 T Using Interleaved 23 Na / 1 H pTx MRI With Motion and Anatomy‐Based B 1 Correction. Magnetic Resonance in Medicine. 2026;96(1):173-190. doi:10.1002/mrm.70342

BibTeX
@article{Ruck2026,
  author = {Ruck, Laurent and Egger, Nico and Zobler, Benedikt and Schirmer, Judith and Nagelstraßer, Sophia and Bitz, Andreas and Platt, Tanja and Konstandin, Simon and Kopp, Christoph and Uder, Michael and Nagel, Armin Michael},
  title = {Improved Myocardial Sodium Quantification at 7 T Using Interleaved 23 Na / 1 H pTx MRI With Motion and Anatomy‐Based B 1 Correction},
  journal = {Magnetic Resonance in Medicine},
  year = {2026},
  volume = {96},
  number = {1},
  pages = {173--190},
  doi = {10.1002/mrm.70342}
}
Start for free →