A bootstrapping residuals approach to determine the error in quantitative functional lung imaging
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30367 ↗At a glance
Journal article in Magnetic Resonance in Medicine, 2025. Written by Anne Slawig and 3 co-authors. Institution: Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Würzburg, Germany. Major MEDLINE topics: Lung, Cystic Fibrosis, Magnetic Resonance Imaging, Image Interpretation, Computer-Assisted. Cited 1 times so far (Crossref); 46 references. The full text is freely available (open access, CC BY 4.0). Funded by: German Research Foundation, Deutsche Forschungsgemeinschaft.
Abstract (original language)
Purpose
To implement and validate an algorithm to determine the statistical errors in self‐gated non‐contrast‐enhanced functional lung imaging.
Methods
A bootstrapping residuals approach to determine the error in quantitative functional lung imaging is proposed. Precision and accuracy of the median error over the lungs, as well as reproducibility of the approach were investigated in 7 volunteers. The algorithm was additionally applied to data acquired in a patient with cystic fibrosis.
Results
The obtained bootstrapping error maps appear comparable to the error maps determined from repeated measurements, and median absolute error values for both methods show comparable median errors when reducing the number of averages. In a volunteer in whom 10 consecutive measurements were carried out, the median functional parameters were ventilation = 0.22 mL gas/mL lung tissue, perfusion amplitude = 0.028, perfusion timing = −82 ms, whereas precision and accuracy of the median error were below 3.2 × 10 −3 mL gas/mL lung for ventilation tissue, 4.4 × 10 −4 for perfusion amplitude, and 11 ms for perfusion timing. In the measurement of the patient, low errors in areas with reduced ventilation support the assessment as real defects.
Conclusion
Using a bootstrapping residuals method, the error of functional lung MRI could be determined without the need for repeated measurements. The error values can be determined reproducibly and can be used as a future means of quality control for functional lung MRI.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Würzburg, Germany.
- University Clinic and Outpatient Clinic for Radiology, Medical Physics Group, University Hospital Halle (Saale), Halle, Germany.
- Halle MR Imaging Core Facility, Medical Faculty, Martin Luther University Halle Wittenberg, Halle, Germany.
- Division of Pediatric Radiology, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
ORCID profiles: Anne Slawig, Andreas Max Weng
Funding
- German Research Foundation
- Deutsche Forschungsgemeinschaft
Cite
APA
Slawig, A., Weng, A. M., Veldhoen, S., & Köstler, H. (2025). A bootstrapping residuals approach to determine the error in quantitative functional lung imaging. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30367
Vancouver
Slawig A, Weng AM, Veldhoen S, Köstler H. A bootstrapping residuals approach to determine the error in quantitative functional lung imaging. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30367
BibTeX
@article{Slawig2025,
author = {Slawig, Anne and Weng, Andreas Max and Veldhoen, Simon and Köstler, Herbert},
title = {A bootstrapping residuals approach to determine the error in quantitative functional lung imaging},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30367}
}
EUVIMED