Feasibility and SNR Performance of Hyperpolarized 129 Xe Gas Exchange Imaging Using a Balanced SSFP Sequence
Magnetic Resonance in Medicine · 2026
DOI: 10.1002/mrm.70503 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2026. Verfasst von Agilo L Kern und 4 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany. Hauptthemen laut MEDLINE: Xenon Isotopes, Magnetic Resonance Imaging, Lung, Image Processing, Computer-Assisted. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: German Center for Lung Research (DZL), Deutsches Zentrum für Lungenforschung.
Abstract (Originalsprache)
Purpose
To develop a balanced SSFP (bSSFP) sequence for pulmonary 129 Xe dissolved‐phase imaging based on a previously described double‐echo sequence facilitating separation into red blood cell and membrane signals, to test its feasibility and compare its SNR performance and imaging results to an analogous FLASH sequence.
Methods
12 healthy volunteers underwent hyperpolarized 129 Xe MRI with both sequences with two different choices of flip angle for FLASH. SNR was quantified using a pseudo multiple replica method, normalized by 129 Xe polarization and ventilated volume. Chemical shift separation was performed in k‐space taking results from time‐domain fits of signal evolution as input, either approximating T E as constant or taking the finite readout duration into account. Results were compared between sequences and methods by linear mixed models, Bland–Altman, and scatter plots.
Results
Normalized SNR of the first echo was significantly greater for bSSFP compared to FLASH with the same flip angle, p = 0.003, but not with lower flip angle resulting in equivalent depolarization, p = 0.98. The effective number of signal averages in signal decomposition was significantly higher compared to FLASH with both flip angles and both dissolved resonances, p < 0.008. Neglecting the finite readout duration was found to introduce bias of derived RBC‐M ratios but prevent noise amplification.
Conclusion
Hyperpolarized 129 Xe dissolved‐phase imaging using dual‐echo bSSFP is feasible and provides superior effective number of signal averages resulting in improved SNR efficiency of signal decomposition. Future work should concentrate on the application in patients with lung disease.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Institute for Diagnostic and Interventional Radiology, Hannover Medical School, Hannover, Germany.
- Biomedical Research in Endstage and Obstructive Lung Disease Hannover (BREATH), German Center for Lung Research (DZL), Hannover, Germany.
- Department of Radiology, Charité - University Medical Center Berlin, Corporate Member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
- Clinical Airway Research, Fraunhofer Institute for Toxicology and Experimental Medicine (ITEM), Hannover, Germany.
- Department of Respiratory Medicine and Infectious Diseases, Hannover Medical School, Hannover, Germany.
ORCID-Profile: Agilo L Kern
Förderung
- German Center for Lung Research (DZL)
- Deutsches Zentrum für Lungenforschung
Zitieren
APA
Kern, A. L., Gutberlet, M., Hohlfeld, J. M., Wacker, F., & Vogel-Claussen, J. (2026). Feasibility and SNR Performance of Hyperpolarized 129 Xe Gas Exchange Imaging Using a Balanced SSFP Sequence. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.70503
Vancouver
Kern AL, Gutberlet M, Hohlfeld JM, Wacker F, Vogel-Claussen J. Feasibility and SNR Performance of Hyperpolarized 129 Xe Gas Exchange Imaging Using a Balanced SSFP Sequence. Magnetic Resonance in Medicine. 2026. doi:10.1002/mrm.70503