EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Analytical corrections for B 1 ‐inhomogeneity and signal decay in multi‐slice 2D spiral hyperpolarized 129 Xe MRI using keyhole reconstruction

J W Plummer iD, R Hussain iD, A S Bdaiwi iD, M L Costa iD, M M Willmering iD, J Parra-Robles iD, Z I Cleveland iD, L L Walkup iD

Magnetic Resonance in Medicine · 2024

DOI: 10.1002/mrm.30028 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2024. Verfasst von J W Plummer und 7 weiteren Autor:innen. Beteiligte Einrichtung: Center for Pulmonary Imaging Research, Division of Pulmonary Medicine, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Xenon Isotopes, Artifacts, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: NHLBI NIH HHS, NIH HHS, Cystic Fibrosis Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Hyperpolarized xenon MRI suffers from heterogeneous coil bias and magnetization decay that obscure pulmonary abnormalities. Non‐physiological signal variability can be mitigated by measuring and mapping the nominal flip angle, and by rescaling the images to correct for signal bias and decay. While flip angle maps can be calculated from sequentially acquired images, scan time and breath‐hold duration are doubled. Here, we exploit the low‐frequency oversampling of 2D‐spiral and keyhole reconstruction to measure flip angle maps from a single acquisition.

Methods

Flip angle maps were calculated from two images generated from a single dataset using keyhole reconstructions and a Bloch‐equation–based model suitable for hyperpolarized substances. Artifacts resulting from acquisition and reconstruction schemes (e.g., keyhole reconstruction radius, slice‐selection profile, spiral‐ordering, and oversampling) were assessed using point‐spread functions. Simulated flip angle maps generated using keyhole reconstruction were compared against the paired‐image approach using RMS error (RMSE). Finally, feasibility was demonstrated for in vivo xenon ventilation imaging.

Results

Simulations demonstrated accurate flip angle maps and B 1 ‐inhomogeneity correction can be generated with only 1.25‐fold central‐oversampling and keyhole reconstruction radius = 5% (RMSE = 0.460°). These settings also generated accurate flip angle maps in a healthy control (RSME = 0.337°) and a person with cystic fibrosis (RMSE = 0.404°) in as little as 3.3 s.

Conclusion

Regional lung ventilation images with reduced impact of B 1 ‐inhomogeneity can be acquired rapidly by combining 2D‐spiral acquisition, Bloch‐equation–based modeling, and keyhole reconstruction. This approach will be especially useful for breath‐hold studies where short scan durations are necessary, such as dynamic imaging and applications in children or people with severely compromised respiratory function.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingXenon IsotopesArtifactsImage Processing, Computer-AssistedLungHumansComputer SimulationAlgorithmsMaleFemalePhantoms, ImagingAdultBreath HoldingCystic Fibrosis

Stichwörter der Autor:innen

2D spiralB1‐inhomogeneity correctionhyperpolarized 129Xekeyhole reconstructionlung imagingmulti‐sliceslice‐selective

Substanzen

Xenon IsotopesXenon-129

Autor:innen und Institutionen

  • Center for Pulmonary Imaging Research, Division of Pulmonary Medicine, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Department of Biomedical Engineering, University of Cincinnati, Cincinnati, OH, USA.
  • Imaging Research Center, Department of Radiology, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA.
  • Department of Pediatrics, University of Cincinnati, Cincinnati, OH, USA.

ORCID-Profile: J W Plummer, R Hussain, A S Bdaiwi, M L Costa, M M Willmering, J Parra-Robles, Z I Cleveland, L L Walkup

Förderung

  • NHLBI NIH HHS
  • NIH HHS
  • Cystic Fibrosis Foundation
  • National Institutes of Health

Zitieren

APA

Plummer, J. W., Hussain, R., Bdaiwi, A. S., Costa, M. L., Willmering, M. M., Parra-Robles, J., Cleveland, Z. I., & Walkup, L. L. (2024). Analytical corrections for B 1 ‐inhomogeneity and signal decay in multi‐slice 2D spiral hyperpolarized 129 Xe MRI using keyhole reconstruction. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30028

Vancouver

Plummer JW, Hussain R, Bdaiwi AS, Costa ML, Willmering MM, Parra-Robles J, et al. Analytical corrections for B 1 ‐inhomogeneity and signal decay in multi‐slice 2D spiral hyperpolarized 129 Xe MRI using keyhole reconstruction. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30028

BibTeX
@article{Plummer2024,
  author = {Plummer, J W and Hussain, R and Bdaiwi, A S and Costa, M L and Willmering, M M and Parra-Robles, J and Cleveland, Z I and Walkup, L L},
  title = {Analytical corrections for B 1 ‐inhomogeneity and signal decay in multi‐slice 2D spiral hyperpolarized 129 Xe MRI using keyhole reconstruction},
  journal = {Magnetic Resonance in Medicine},
  year = {2024},
  doi = {10.1002/mrm.30028}
}
Kostenlos starten →