Ultra‐low‐field magnetization transfer imaging at 0.055T with low specific absorption rate
Magnetic Resonance in Medicine · 2024-07-24
DOI: 10.1002/mrm.30231 ↗At a glance
Journal article in Magnetic Resonance in Medicine, published online on 2024-07-24. Written by Shi Su and 10 co-authors. Institution: Laboratory of Biomedical Imaging and Signal Processing, The University of Hong Kong, Hong Kong SAR, China. Major MEDLINE topics: Magnetic Resonance Imaging, Phantoms, Imaging, Brain. Cited 7 times so far (Crossref); 51 references. The full text is freely available (open access, CC BY 4.0). Funded by: Hong Kong Research Grant Council.
Abstract (original language)
Purpose
To demonstrate magnetization transfer (MT) effects with low specific absorption rate (SAR) on ultra‐low‐field (ULF) MRI.
Methods
MT imaging was implemented by using sinc‐modulated RF pulse train (SPT) modules to provide bilateral off‐resonance irradiation. They were incorporated into 3D gradient echo (GRE) and fast spin echo (FSE) protocols on a shielding‐free 0.055T head scanner. MT effects were first verified using phantoms. Brain MT imaging was conducted in both healthy subjects and patients.
Results
MT effects were clearly observed in phantoms using six SPT modules with total flip angle 3600° at central primary saturation bands of approximate offset ±786 Hz, even in the presence of large relative B 0 inhomogeneity. For brain, strong MT effects were observed in gray matter, white matter, and muscle in 3D GRE and FSE imaging using six and sixteen SPT modules with total flip angle 3600° and 9600°, respectively. Fat, cerebrospinal fluid, and blood exhibited relatively weak MT effects. MT preparation enhanced tissue contrasts in T2‐weighted and FLAIR‐like images, and improved brain lesion delineation. The estimated MT SAR was 0.0024 and 0.0008 W/kg for two protocols, respectively, which is far below the US Food and Drug Administration (FDA) limit of 3.0 W/kg.
Conclusion
Robust MT effects can be readily obtained at ULF with extremely low SAR, despite poor relative B 0 homogeneity in ppm. This unique advantage enables flexible MT pulse design and implementation on low‐cost ULF MRI platforms to achieve strong MT effects in brain and beyond, potentially augmenting their clinical utility in the future.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Laboratory of Biomedical Imaging and Signal Processing, The University of Hong Kong, Hong Kong SAR, China.
- Department of Electrical and Electronic Engineering, The University of Hong Kong, Hong Kong SAR, China.
- Department of Surgery, The University of Hong Kong, Hong Kong SAR, China.
- Department of Medicine, The University of Hong Kong, Hong Kong SAR, China.
- Department of Diagnostic Radiology, The University of Hong Kong, Hong Kong SAR, China.
Funding
- Hong Kong Research Grant Council
Cite
APA
Su, S., Zhao, Y., Ding, Y., Lau, V., Xiao, L., Leung, G. K. K., Lau, G. K. K., Huang, F., Vardhanabhuti, V., Leong, A. T. L., & Wu, E. X. (2024). Ultra‐low‐field magnetization transfer imaging at 0.055T with low specific absorption rate. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30231
Vancouver
Su S, Zhao Y, Ding Y, Lau V, Xiao L, Leung GKK, et al. Ultra‐low‐field magnetization transfer imaging at 0.055T with low specific absorption rate. Magnetic Resonance in Medicine. 2024. doi:10.1002/mrm.30231
BibTeX
@article{Su2024,
author = {Su, Shi and Zhao, Yujiao and Ding, Ye and Lau, Vick and Xiao, Linfang and Leung, Gilberto K K and Lau, Gary K K and Huang, Fan and Vardhanabhuti, Vince and Leong, Alex T L and Wu, Ed X},
title = {Ultra‐low‐field magnetization transfer imaging at 0.055T with low specific absorption rate},
journal = {Magnetic Resonance in Medicine},
year = {2024},
doi = {10.1002/mrm.30231}
}
EUVIMED