Time‐efficient, high‐resolution 3T whole‐brain relaxometry using Cartesian 3D MR Spin TomogrAphy in Time‐Domain (MR‐STAT) with cerebrospinal fluid suppression
Magnetic Resonance in Medicine · 2025
DOI: 10.1002/mrm.30384 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, 2025. Verfasst von Hongyan Liu und 6 weiteren Autor:innen. Beteiligte Einrichtung: Computational Imaging Group, Department of Radiotheraphy, University Medical Center Utrecht, Utrecht, The Netherlands. Hauptthemen laut MEDLINE: Brain, Magnetic Resonance Imaging, Imaging, Three-Dimensional, Cerebrospinal Fluid. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: China Scholarship Council.
Abstract (Originalsprache)
Purpose
Current three‐dimensional (3D) MR Spin TomogrAphy in Time‐Domain (MR‐STAT) protocols use transient‐state, gradient‐spoiled gradient‐echo sequences that are prone to cerebrospinal fluid (CSF) pulsation artifacts when applied to the brain. This study aims to develop a 3D MR‐STAT protocol for whole‐brain relaxometry that overcomes the challenges posed by CSF‐induced ghosting artifacts.
Method
We optimized the flip‐angle train within the Cartesian 3D MR‐STAT framework to achieve two objectives: (1) minimization of the noise level in the reconstructed quantitative maps, and (2) reduction of the CSF–to–white‐matter signal ratio to suppress CSF‐associated pulsation artifacts. The optimized new sequence was tested on a gel/water phantom for accuracy evaluation of the quantitative maps, and on healthy volunteers to explore the effectiveness of the CSF artifact suppression and robustness of the new protocol.
Results
An optimized sequence with high parameter‐encoding capability and low CSF signal response was proposed and validated in the gel/water phantom experiment. From in vivo experiments with 5 volunteers, the proposed CSF‐suppressed sequence produced quantitative maps with no CSF artifacts and showed overall greatly improved image quality compared with the baseline sequence. Statistical analysis indicated low intersubject and interscan variability for quantitative parameters in gray matter and white matter (1.6%–2.4% for T 1 and 2.0%–4.6% for T 2 ), demonstrating the robustness of the new sequence.
Conclusion
We present a new 3D MR‐STAT sequence with CSF suppression that effectively eliminates CSF pulsation artifacts. The new sequence ensures consistently high‐quality, 1‐mm 3 whole‐brain relaxometry within a rapid 5.5‐min scan time.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Computational Imaging Group, Department of Radiotheraphy, University Medical Center Utrecht, Utrecht, The Netherlands.
ORCID-Profile: Hongyan Liu, Edwin Versteeg, Miha Fuderer, Oscar van der Heide, Martin B Schilder, Cornelis A T van den Berg, Alessandro Sbrizzi
Förderung
- China Scholarship Council
Zitieren
APA
Liu, H., Versteeg, E., Fuderer, M., van der Heide, O., Schilder, M. B., van den Berg, C. A. T., & Sbrizzi, A. (2025). Time‐efficient, high‐resolution 3T whole‐brain relaxometry using Cartesian 3D MR Spin TomogrAphy in Time‐Domain (MR‐STAT) with cerebrospinal fluid suppression. Magnetic Resonance in Medicine. https://doi.org/10.1002/mrm.30384
Vancouver
Liu H, Versteeg E, Fuderer M, van der Heide O, Schilder MB, van den Berg CAT, et al. Time‐efficient, high‐resolution 3T whole‐brain relaxometry using Cartesian 3D MR Spin TomogrAphy in Time‐Domain (MR‐STAT) with cerebrospinal fluid suppression. Magnetic Resonance in Medicine. 2025. doi:10.1002/mrm.30384
BibTeX
@article{Liu2025,
author = {Liu, Hongyan and Versteeg, Edwin and Fuderer, Miha and van der Heide, Oscar and Schilder, Martin B and van den Berg, Cornelis A T and Sbrizzi, Alessandro},
title = {Time‐efficient, high‐resolution 3T whole‐brain relaxometry using Cartesian 3D MR Spin TomogrAphy in Time‐Domain (MR‐STAT) with cerebrospinal fluid suppression},
journal = {Magnetic Resonance in Medicine},
year = {2025},
doi = {10.1002/mrm.30384}
}
EUVIMED