Rapid quantitative MRI at 46 mT : Accelerated T 1 and T 2 mapping with low‐rank reconstructions
Magnetic Resonance in Medicine · 2025-07 · 94 (1) · 119-133
DOI: 10.1002/mrm.30442 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 29.01.2025. Verfasst von Yiming Dong und 5 weiteren Autor:innen. Beteiligte Einrichtung: C.J. Gorter MRI Center, Department of Radiology, Leiden University Medical Center (LUMC), Leiden, The Netherlands. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Image Processing, Computer-Assisted, Brain. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Dutch Science Foundation (NWO) Open Technology, Nederlandse Organisatie voor Wetenschappelijk Onderzoek.
Abstract (Originalsprache)
Purpose
To evaluate accelerated T 1 ‐ and T 2 ‐mapping techniques for ultra–low‐field MRI using low‐rank reconstruction methods.
Methods
Two low‐rank–based algorithms, image‐based locally low‐rank (LLR) and k‐space–based structured low‐rank (SLR), were implemented to accelerate T 1 and T 2 mapping on a 46 mT Halbach MRI scanner. Data were acquired with 3D turbo spin‐echo sequences using variable‐density poisson‐disk random sampling patterns. For validation, phantom and in vivo experiments were performed on six healthy volunteers to compare the obtained values with literature and to study reconstruction performance at different undersampling factors and spatial resolutions. In addition, the reconstruction performance of the LLR and SLR algorithms for T 1 mapping was compared using retrospective undersampling datasets. Total scan times were reduced from 45/38 min (R = 1) to 23/19 min (R = 2) and 11/9 min (R = 4) for a 2.5 × 2.5 × 5 mm 3 resolution, and to 18/16 min (R = 4) for a higher in‐plane resolution 1.5 × 1.5 × 5 mm 3 for T 1 /T 2 mapping, respectively.
Results
Both LLR and SLR algorithms successfully reconstructed T 1 and T 2 maps from undersampled data, significantly reducing scan times and eliminating undersampling artifacts. Phantom validation showed that consistent T 1 and T 2 values were obtained at different undersampling factors up to R = 4. For in vivo experiments, comparable image quality and estimated T 1 and T 2 values were obtained for fully sampled and undersampled (R = 4) reconstructions, both of which were in line with the literature values.
Conclusions
The use of low‐rank reconstruction allows significant acceleration of T 1 and T 2 mapping in low‐field MRI while maintaining image quality.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- C.J. Gorter MRI Center, Department of Radiology, Leiden University Medical Center (LUMC), Leiden, The Netherlands.
- Philips Research Hamburg, Hamburg, Germany.
ORCID-Profile: Yiming Dong, Chloé Najac, Matthias J P van Osch, Beatrice Lena
Förderung
- Dutch Science Foundation (NWO) Open Technology
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek
Zitieren
APA
Dong, Y., Najac, C., van Osch, M. J. P., Webb, A., Börnert, P., & Lena, B. (2025). Rapid quantitative MRI at 46 mT : Accelerated T 1 and T 2 mapping with low‐rank reconstructions. Magnetic Resonance in Medicine, 94(1), 119-133. https://doi.org/10.1002/mrm.30442
Vancouver
Dong Y, Najac C, van Osch MJP, Webb A, Börnert P, Lena B. Rapid quantitative MRI at 46 mT : Accelerated T 1 and T 2 mapping with low‐rank reconstructions. Magnetic Resonance in Medicine. 2025;94(1):119-133. doi:10.1002/mrm.30442
BibTeX
@article{Dong2025,
author = {Dong, Yiming and Najac, Chloé and van Osch, Matthias J P and Webb, Andrew and Börnert, Peter and Lena, Beatrice},
title = {Rapid quantitative MRI at 46 mT : Accelerated T 1 and T 2 mapping with low‐rank reconstructions},
journal = {Magnetic Resonance in Medicine},
year = {2025},
volume = {94},
number = {1},
pages = {119--133},
doi = {10.1002/mrm.30442}
}