Development and evaluation of an in vivo dose‐based monitoring system for electron FLASH radiation therapy

Justin DeFrancisco, Matthew Richeson, Tomaj Javidtash, Chris Bartee, Tianjun Ma, Siyong Kim

Medical Physics · 2026-07 · 53 (7)

DOI: 10.1002/mp.70549 ↗

At a glance

Evaluationsstudie in Medical Physics, published online on 2026-06-30. Written by Justin DeFrancisco and 5 co-authors. Institution: Department of Radiation Oncology, School of Medicine, Virginia Commonwealth University Health System, Richmond, Virginia, USA. Major MEDLINE topics: Electrons, Radiation Dosage, Radiometry, Radiotherapy. 53 references. The full text is freely available (open access, CC BY 4.0). Funded by: VCU Intellectual property foundation.

AbstraktSubjectsDane bibliograficzneCiteRelated studies

Abstract (original language)

Background

FLASH radiotherapy requires further preclinical and clinical investigation to establish its biological effectiveness and define optimal beam parameters. In conventional (CONV) radiotherapy, redundant beam termination systems are a cornerstone ensuring patient safety, yet analogous safeguards for FLASH delivery are not well established, creating a critical barrier to safely enabling such studies.

Purpose

To develop and evaluate a real‐time, in vivo, point‐dose monitoring system capable of terminating electron FLASH beam delivery as an additional monitoring system on a modified medical linear accelerator (LINAC).

Methods

A decommissioned LINAC was modified to deliver electron FLASH beams with stable dose per pulse (DPP) at 300 Hz pulse frequency. A commercial plastic scintillation detector system was adapted through hardware and firmware modifications to enable pulse‐based and dose‐based beam termination via the LINAC MLC interface. The detector was cross‐calibrated against radiochromic film under FLASH conditions. System performance was evaluated through measurements of control accuracy, and detector response as a function of DPP.

Results

Stable electron FLASH delivery was achieved with an average dose rate of 127.514.91 Gy/s and an approximate beam energy of 5.2 MeV. Pulse‐based control terminated delivery within +3 pulses of the requested value (requested 1‐20 pulses), with overshoot attributable to downstream circuitry latency. Dose‐based control agreed with film measurements within 1.110.81 Gy for surface‐based control (in vivo setup) and −1.450.38 Gy at depth (stable dosimetry) (tested dose deliveries between 2‐15 Gy). The detector response versus DPP in the FLASH range (0.11–0.78 Gy/p) could be roughly approximated as linear before detector saturation, with only marginal improvement seen when using quadratic fitting.

Conclusion

A modified scintillation‐based system was implemented as a real‐time in vivo beam termination mechanism for electron FLASH radiotherapy under stable DPP and specific experimental conditions. While not intended for primary beam control, the system may provide a practical redundant safety layer for mitigating gross delivery errors in experimental and translational FLASH applications.

Subjects

MeSH terms (MEDLINE)

ElectronsRadiation DosageRadiometryRadiotherapyRadiotherapy DosageParticle AcceleratorsScintillation CountingAnimals

Author keywords

Flashin vivoscintillator

Authors and institutions

  • Department of Radiation Oncology, School of Medicine, Virginia Commonwealth University Health System, Richmond, Virginia, USA.
  • Medical Physics Department, Virginia Commonwealth University, Richmond, Virginia, USA.

Funding

  • VCU Intellectual property foundation

Cite

APA

DeFrancisco, J., Richeson, M., Javidtash, T., Bartee, C., Ma, T., & Kim, S. (2026). Development and evaluation of an in vivo dose‐based monitoring system for electron FLASH radiation therapy. Medical Physics, 53(7). https://doi.org/10.1002/mp.70549

Vancouver

DeFrancisco J, Richeson M, Javidtash T, Bartee C, Ma T, Kim S. Development and evaluation of an in vivo dose‐based monitoring system for electron FLASH radiation therapy. Medical Physics. 2026;53(7). doi:10.1002/mp.70549

BibTeX
@article{DeFrancisco2026,
  author = {DeFrancisco, Justin and Richeson, Matthew and Javidtash, Tomaj and Bartee, Chris and Ma, Tianjun and Kim, Siyong},
  title = {Development and evaluation of an in vivo dose‐based monitoring system for electron FLASH radiation therapy},
  journal = {Medical Physics},
  year = {2026},
  volume = {53},
  number = {7},
  doi = {10.1002/mp.70549}
}
Start for free →