Dosimetric calibration of anatomy‐specific ultra‐high dose rate electron irradiation platform for preclinical FLASH radiobiology experiments

Jinghui Wang, Stavros Melemenidis, Rakesh Manjappa, Vignesh Viswanathan, Ramish M Ashraf, Karen Levy, Lawrie B Skinner, Luis A Soto, Stephanie Chow, Brianna Lau, Ryan B Ko, Edward E Graves und 7 weitere

Medical Physics · 2024

DOI: 10.1002/mp.17432 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Jinghui Wang und 18 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Stanford University School of Medicine, Stanford, California, USA. Hauptthemen laut MEDLINE: Electrons, Radiometry, Radiobiology, Radiotherapy Dosage. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

FLASH radiation therapy (RT) offers a promising avenue for the broadening of the therapeutic index. However, to leverage the full potential of FLASH in the clinical setting, an improved understanding of the biological principles involved is critical. This requires the availability of specialized equipment optimized for the delivery of conventional (CONV) and ultra‐high dose rate (UHDR) irradiation for preclinical studies. One method to conduct such preclinical radiobiological research involves adapting a clinical linear accelerator configured to deliver both CONV and UHDR irradiation.

Purpose

We characterized the dosimetric properties of a clinical linear accelerator configured to deliver ultra‐high dose rate irradiation to two anatomic sites in mice and for cell‐culture FLASH radiobiology experiments.

Methods

Delivered doses of UHDR electron beams were controlled by a microcontroller and relay interfaced with the respiratory gating system. We also produced beam collimators with indexed stereotactic mouse positioning devices to provide anatomically specific preclinical treatments. Treatment delivery was monitored directly with an ionization chamber, and charge measurements were correlated with radiochromic film measurements at the entry surface of the mice. The setup for conventional dose rate irradiation utilized the same collimation system but at increased source‐to‐surface distance. Monte Carlo simulations and film dosimetry were used to characterize beam properties and dose distributions.

Results

The mean electron beam energies before the flattening filter were 18.8 MeV (UHDR) and 17.7 MeV (CONV), with corresponding values at the mouse surface of 17.2 and 16.2 MeV. The charges measured with an external ion chamber were linearly correlated with the mouse entrance dose. The use of relay gating for pulse control initially led to a delivery failure rate of 20% (± 1 pulse); adjustments to account for the linac latency improved this rate to < 1/20. Beam field sizes for two anatomically specific mouse collimators (4 × 4 cm 2 for whole‐abdomen and 1.5 × 1.5 cm 2 for unilateral lung irradiation) were accurate within < 5% and had low radiation leakage (< 4%). Normalizing the dose at the center of the mouse (∼0.75 cm depth) produced UHDR and CONV doses to the irradiated volumes with > 95% agreement.

Conclusion

We successfully configured a clinical linear accelerator for increased output and developed a robust preclinical platform for anatomically specific irradiation, with highly accurate and precise temporal and spatial dose delivery, for both CONV and UHDR irradiation applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectronsRadiometryRadiobiologyRadiotherapy DosageAnimalsMiceCalibrationMonte Carlo MethodParticle AcceleratorsRadiation Dosage

Stichwörter der Autor:innen

FLASHLinacdosimetryradiotherapyultra‐high dose rate

Autor:innen und Institutionen

  • Department of Radiation Oncology, Stanford University School of Medicine, Stanford, California, USA.
  • Department of Gynecologic Oncology, Stanford University School of Medicine, Stanford, California, USA.
  • Stanford Cancer Institute, Stanford University School of Medicine, Stanford, California, USA.

Förderung

  • NCI NIH HHS

Zitieren

APA

Wang, J., Melemenidis, S., Manjappa, R., Viswanathan, V., Ashraf, R. M., Levy, K., Skinner, L. B., Soto, L. A., Chow, S., Lau, B., Ko, R. B., Graves, E. E., Yu, A. S., Bush, K. K., Surucu, M., Rankin, E. B., Loo, B. W., Schüler, E., & Maxim, P. G. (2024). Dosimetric calibration of anatomy‐specific ultra‐high dose rate electron irradiation platform for preclinical FLASH radiobiology experiments. Medical Physics. https://doi.org/10.1002/mp.17432

Vancouver

Wang J, Melemenidis S, Manjappa R, Viswanathan V, Ashraf RM, Levy K, et al. Dosimetric calibration of anatomy‐specific ultra‐high dose rate electron irradiation platform for preclinical FLASH radiobiology experiments. Medical Physics. 2024. doi:10.1002/mp.17432

BibTeX
@article{Wang2024,
  author = {Wang, Jinghui and Melemenidis, Stavros and Manjappa, Rakesh and Viswanathan, Vignesh and Ashraf, Ramish M and Levy, Karen and Skinner, Lawrie B and Soto, Luis A and Chow, Stephanie and Lau, Brianna and Ko, Ryan B and Graves, Edward E and Yu, Amy S and Bush, Karl K and Surucu, Murat and Rankin, Erinn B and Loo, Billy W and Schüler, Emil and Maxim, Peter G},
  title = {Dosimetric calibration of anatomy‐specific ultra‐high dose rate electron irradiation platform for preclinical FLASH radiobiology experiments},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.17432}
}
Kostenlos starten →