A comprehensive investigation of the performance of a commercial scintillator system for applications in electron FLASH radiotherapy

Kevin Liu, Shannon Holmes, Emil Schüler, Sam Beddar

Medical Physics · 2024-06 · 51 (6) · 4504-4512

DOI: 10.1002/mp.17030 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 20.03.2024. Verfasst von Kevin Liu und 3 weiteren Autor:innen. Beteiligte Einrichtung: Division of Radiation Oncology, Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Electrons, Scintillation Counting. Die Arbeit wurde bisher 24-mal zitiert (Crossref) und nennt 57 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, Cancer Prevention and Research Institute of Texas, National Cancer Institute of the National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Dosimetry in ultra‐high dose rate (UHDR) beamlines is significantly challenged by limitations in real‐time monitoring and accurate measurement of beam output, beam parameters, and delivered doses using conventional radiation detectors, which exhibit dependencies in ultra‐high dose‐rate (UHDR) and high dose‐per‐pulse (DPP) beamline conditions.

Purpose

In this study, we characterized the response of the Exradin W2 plastic scintillator (Standard Imaging, Inc.), a water‐equivalent detector that provides measurements with a time resolution of 100 Hz, to determine its feasibility for use in UHDR electron beamlines.

Methods

The W2 scintillator was exposed to an UHDR electron beam with different beam parameters by varying the pulse repetition frequency (PRF), pulse width (PW), and pulse amplitude settings of an electron UHDR linear accelerator system. The response of the W2 scintillator was evaluated as a function of the total integrated dose delivered, DPP, and mean and instantaneous dose rate. To account for detector radiation damage, the signal sensitivity (pC/Gy) of the W2 scintillator was measured and tracked as a function of dose history.

Results

The W2 scintillator demonstrated mean dose rate independence and linearity as a function of integrated dose and DPP for DPP ≤ 1.5 Gy ( R 2 > 0.99) and PRF ≤ 90 Hz. At DPP > 1.5 Gy, nonlinear behavior and signal saturation in the blue and green signals as a function of DPP, PRF, and integrated dose became apparent. In the absence of Cerenkov correction, the W2 scintillator exhibited PW dependence, even at DPP values 1.5 Gy, the W2 scintillator's response becomes nonlinear, likely due to limitations in the electrometer related to the high signal intensity.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectronsScintillation CountingRadiotherapy DosageRadiometryRadiotherapyParticle Accelerators

Stichwörter der Autor:innen

FLASHUHDRW2dosimetryplastic scintillatorscintillator

Autor:innen und Institutionen

  • Division of Radiation Oncology, Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • UTHealth Houston Graduate School of Biomedical Sciences, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Standard Imaging Inc., Middleton, Wisconsin, USA.

Förderung

  • NCI NIH HHS
  • Cancer Prevention and Research Institute of Texas
  • National Cancer Institute of the National Institutes of Health

Zitieren

APA

Liu, K., Holmes, S., Schüler, E., & Beddar, S. (2024). A comprehensive investigation of the performance of a commercial scintillator system for applications in electron FLASH radiotherapy. Medical Physics, 51(6), 4504-4512. https://doi.org/10.1002/mp.17030

Vancouver

Liu K, Holmes S, Schüler E, Beddar S. A comprehensive investigation of the performance of a commercial scintillator system for applications in electron FLASH radiotherapy. Medical Physics. 2024;51(6):4504-4512. doi:10.1002/mp.17030

BibTeX
@article{Liu2024,
  author = {Liu, Kevin and Holmes, Shannon and Schüler, Emil and Beddar, Sam},
  title = {A comprehensive investigation of the performance of a commercial scintillator system for applications in electron FLASH radiotherapy},
  journal = {Medical Physics},
  year = {2024},
  volume = {51},
  number = {6},
  pages = {4504--4512},
  doi = {10.1002/mp.17030}
}
Kostenlos starten →