Quantifying multi‐institutional ADC measurement variability of 1.5 T MR‐Linacs: A phantom and in vivo study

Madeline E Carr, Kathryn E Keenan, Michaela Beavan, Hilary Byrne, Satomi Higuchi, Amy Walker, Sarah Elliott, John Baines, Vikneswary Batumalai, Peter Metcalfe, Lois Holloway, Michael G Jameson

Medical Physics · 2025

DOI: 10.1002/mp.17739 ↗

At a glance

Multizentrische Studie in Medical Physics, 2025. Written by Madeline E Carr and 11 co-authors. Institution: GenesisCare, Sydney, New South Wales, Australia. Major MEDLINE topics: Phantoms, Imaging, Diffusion Magnetic Resonance Imaging, Particle Accelerators. Cited 3 times so far (Crossref); 27 references. The full text is freely available (open access, CC BY 4.0).

ResumenSubjectsDatos bibliográficosCiteRelated studies

Abstract (original language)

Background

Diffusion‐weighted imaging (DWI), a quantitative magnetic resonance imaging (qMRI) technique, has the potential to aid in disease characterization and treatment response monitoring. MR‐Linacs (MRLs) enable simultaneous DWI acquisitions during radiotherapy, uniquely aiding in the collection of large‐scale datasets for imaging biomarkers, such as the DWI‐derived apparent diffusion coefficient (ADC), without additional patient burden. However, the limited data reporting on variability in MRL scanner performance characteristics, and a lack of established clinical trial quality assurance (QA) procedures, are barriers to this route for biomarker validation.

Purpose

This study aims to quantify the accuracy, intra‐scanner repeatability, and inter‐scanner reproducibility of ADC measurements across three MRLs in Australia in both a phantom and in vivo. These measurements will inform the feasibility of carrying out prospective multi‐center studies in Australia investigating ADC as a biomarker and form a core set of QA procedures and baselines to assess biomarker and sequence suitability.

Methods

An isotropic diffusion phantom (at 0°C) and one healthy volunteer were scanned on three Unity MRLs (Elekta AB, Stockholm, Sweden). Standardized (QIBA Diffusion Profile) and anatomy‐specific DWI sequences, including sequences recommended by the MR‐Linac Consortium Imaging Biomarker Working Group, were used to image the phantom and volunteer. ADC maps generated using the MRL scanner software (inline ADC) and diffusion‐weighted ( b ‐value) images were exported from the scanner console. The latter was used to generate ADC maps using commercial software (offline ADC) for a separate comparative analysis. Performance metrics were computed for each sequence, including a coefficient of variation to assess between‐session intra‐scanner repeatability (CV BS ) and inter‐scanner reproducibility (CV), for each phantom vial and contoured organ. Additionally, using the phantoms’ known ADC vial values, a percentage bias (bias) was calculated to determine ADC accuracy.

Results

Phantom‐based measurements for the standardized QIBA sequence had intra‐ and inter‐scanner CV and bias well within recommended guideline (QIBA Diffusion Profile) tolerance limits of 2.2% and ±3.6%, respectively. All anatomy‐specific phantom DWI sequences were also within these tolerances, except for the cervix sequence at one site which showed an average intra‐scanner bias of +4.5%. Both accuracy and reproducibility for all sequences were worse for lower diffusivity vials measured in the phantom. Additionally, inline and offline ADC maps had high similarity with average percent differences of +0.2%. Volunteer‐based results had worse reproducibility, with the average inter‐scanner CV for the brain and pancreas sequences within 9.0%, however, reaching up to 27.1% for pelvis and abdomen sequences.

Conclusions

This study demonstrated accuracy, intra‐scanner repeatability, and inter‐scanner reproducibility comparable to metrics reported in the literature, using both the phantom and volunteer datasets. The cervix sequence had the largest variability in both phantom and volunteer results and was recommended for further investigation. This study suggests that qMRI techniques utilizing DWI could be a viable option for future multi‐centered patient‐based studies utilizing Australian MRLs, with phantom‐based quality assurance recommended alongside patient imaging.

Subjects

MeSH terms (MEDLINE)

Phantoms, ImagingDiffusion Magnetic Resonance ImagingParticle AcceleratorsHumansReproducibility of ResultsImage Processing, Computer-Assisted

Author keywords

MR‐Linacaccuracyapparent diffusion coefficient (ADC)quantitative magnetic resonance imaging (qMRI)repeatabilityreproducibility

Authors and institutions

  • GenesisCare, Sydney, New South Wales, Australia.
  • Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
  • Liverpool and Macarthur Cancer Therapy Centres/Ingham Institute for Applied Medical Research, Liverpool, Australia.
  • National Institute of Standards and Technology, Boulder, Colorado, USA.
  • School of Clinical Medicine, Medicine and Health, University of New South Wales, Sydney, New South Wales, Australia.
  • School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, New South Wales, Australia.

Cite

APA

Carr, M. E., Keenan, K. E., Beavan, M., Byrne, H., Higuchi, S., Walker, A., Elliott, S., Baines, J., Batumalai, V., Metcalfe, P., Holloway, L., & Jameson, M. G. (2025). Quantifying multi‐institutional ADC measurement variability of 1.5 T MR‐Linacs: A phantom and in vivo study. Medical Physics. https://doi.org/10.1002/mp.17739

Vancouver

Carr ME, Keenan KE, Beavan M, Byrne H, Higuchi S, Walker A, et al. Quantifying multi‐institutional ADC measurement variability of 1.5 T MR‐Linacs: A phantom and in vivo study. Medical Physics. 2025. doi:10.1002/mp.17739

BibTeX
@article{Carr2025,
  author = {Carr, Madeline E and Keenan, Kathryn E and Beavan, Michaela and Byrne, Hilary and Higuchi, Satomi and Walker, Amy and Elliott, Sarah and Baines, John and Batumalai, Vikneswary and Metcalfe, Peter and Holloway, Lois and Jameson, Michael G},
  title = {Quantifying multi‐institutional ADC measurement variability of 1.5 T MR‐Linacs: A phantom and in vivo study},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.17739}
}
Start for free →