EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Geometric distortion in computed tomography scans for radiotherapy: Quantification and quality assurance implications

Yunfei Hu, Emma Cai, John Shakeshaft, Marius Arnesen, Tina Gorjiara, Mikel Byrne, Tim Markwell, Ben Archidbald-Heeren

Journal of Applied Clinical Medical Physics · 2026-07 · 27 (7)

DOI: 10.1002/acm2.70686 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 07.07.2026. Verfasst von Yunfei Hu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Icon Cancer Centre Gosford, Gosford, New South Wales, Australia. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Quality Assurance, Health Care, Cone-Beam Computed Tomography. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Computed tomography (CT) is widely assumed to be geometrically accurate and is commonly used as the reference standard for quantifying geometric distortion in other imaging modalities, particularly magnetic resonance imaging (MRI). However, geometric distortion in CT has not been comprehensively investigated.

Purpose

The aim of this study was to characterize the performance of a distortion phantom, establish a reference scan and, on this basis, determine and quantify the presence of geometric distortion in CT scans used for radiotherapy (RT) planning.

Methods

A CIRS Model 603A Distortion Phantom was employed to quantify geometric distortion. System reproducibility, as well as the effects of phantom setup, scan parameters, fabrication uncertainty, and temporal variation were systematically evaluated. Additionally, the geometric distortion of CT was compared to that of linac‐based cone‐beam CT (CBCT). Subsequently, geometric distortion magnitudes from eight CT scanners were retrospectively analyzed against both CBCT and a Computer Aided Design (CAD) model.

Results

CBCT demonstrated consistently smaller distortion variation compared with CT. Retrospective analysis revealed that all eight analyzed CT scanners exhibited measurable distortion magnitudes, with mean magnitudes ranging from 0.13 mm to 0.52 mm and maximum distortion magnitudes ranging from 0.45 mm to 1.09 mm when compared to CBCT, or 0.25 mm to 0.82 mm and 0.58 mm to 1.88 mm when compared to the CAD model. Longitudinal couch inaccuracy or non‐linear couch sag may constitute the primary contributors to geometric distortion on CT scanners.

Conclusions

Contrary to common assumptions, CT scans used for radiotherapy planning exhibit measurable geometric distortion across clinically relevant fields of view (FoVs). However, current CT QA methodologies and tolerances may lack sufficient sensitivity to detect submillimeter distortion levels relevant to stereotactic radiotherapy (SRT). Recognition and rigorous quantification of CT geometric distortion are essential to inform risk assessment and margin design in high‐precision RT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiotherapy Planning, Computer-AssistedQuality Assurance, Health CareCone-Beam Computed TomographyImage Processing, Computer-AssistedTomography, X-Ray ComputedHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedRetrospective StudiesParticle AcceleratorsReproducibility of Results

Stichwörter der Autor:innen

CTMRIgeometric distortionstereotactic radiotherapy

Autor:innen und Institutionen

  • Icon Cancer Centre Gosford, Gosford, New South Wales, Australia.
  • University of Wollongong, Wollongong, New South Wales, Australia.
  • Icon Cancer Centre Concord, Concord Repatriation General Hospital, Concord, New South Wales, Australia.
  • Icon Cancer Centre Gold Coast, Southport, Queensland, Australia.
  • Icon Cancer Centre Toowoomba, Rockville, Queensland, Australia.
  • Icon Cancer Centre Norwest, Bella Vista, New South Wales, Australia.

Zitieren

APA

Hu, Y., Cai, E., Shakeshaft, J., Arnesen, M., Gorjiara, T., Byrne, M., Markwell, T., & Archidbald-Heeren, B. (2026). Geometric distortion in computed tomography scans for radiotherapy: Quantification and quality assurance implications. Journal of Applied Clinical Medical Physics, 27(7). https://doi.org/10.1002/acm2.70686

Vancouver

Hu Y, Cai E, Shakeshaft J, Arnesen M, Gorjiara T, Byrne M, et al. Geometric distortion in computed tomography scans for radiotherapy: Quantification and quality assurance implications. Journal of Applied Clinical Medical Physics. 2026;27(7). doi:10.1002/acm2.70686

BibTeX
@article{Hu2026,
  author = {Hu, Yunfei and Cai, Emma and Shakeshaft, John and Arnesen, Marius and Gorjiara, Tina and Byrne, Mikel and Markwell, Tim and Archidbald-Heeren, Ben},
  title = {Geometric distortion in computed tomography scans for radiotherapy: Quantification and quality assurance implications},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {7},
  doi = {10.1002/acm2.70686}
}
Kostenlos starten →